16
Pr
OBE
mA
iNTENANCE
CAL Check™ & Cleaning (HI9813-61)
To check meter’s calibration status
:
1. Rinse the probe with water
2. Immerse the tip of the probe 4 cm (1
1/2
”) into
HI50021
calibration
check solution.
Note:
A reading around pH 4.01 indicates that the probe is broken and
requires replacement.
3. Press the
Check
key.
“Probe is OK” is displayed if the meter is calibrated.
“Clean Probe and Calibrate” indicates that cleaning is required.
Screens from DY9813V3 received from Mech.
Dip probe in
solution and turn
the knob to match
calibration
the solution value
pH
Dip probe in
solution and turn
the knob to match
calibration
the solution value
pH
Screens from DY9813V3 received from Mech.
To clean the probe:
1. Immerse the tip of the probe 4 cm (1
1/2
”) into
HI700661
cleaning
solution and soak for 5 minutes.
2. Rinse with tap water.
3. Immerse the tip of the probe in
HI50021
calibration check solution.
4. Press the
Check
key.
“Probe is OK” is displayed if the meter does not require calibration.
“Clean Probe and Calibrate” indicates that the meter requires
calibration. Follow pH Calibration steps.