7. MEASUREMENT
7.1. pH
• Immerse the probe tip 4 cm (1
1/2
”) into the sample.
• Press
pH
key to select pH measurement mode.
• Stir briefly and wait a few minutes for the measurement to stabilize.
Measured pH value is displayed on the screen.
solution and turn the knob
Dip probe in calibration
pH
for current temperature
Screens from DY9812V3 received from Mech.
• If measurements are taken in different samples successively, rinse
the probe thoroughly to minimize cross-contamination. After rinsing
with water, rinse the probe with some of the sample to be measured
next.
7.2. EC / TDS
• Immerse the probe tip 4 cm (1 ½”) into the sample.
If possible, use plastic beakers to minimize any EMC interferences.
• Tap the probe lightly on the bottom of the beaker to remove any air
bubbles trapped inside.
• Press the
μS/cm
or
ppm
key to select EC or TDS measurement mode.
or
• Wait a few minutes. Displayed measured value (in μS/cm or ppm)
is temperature compensated automatically.
solution and turn the knob
Dip probe in calibration
EC/TDS
for current temperature
Screens from DY9812V3 received from Mech.
solution and turn the knob
Dip probe in calibration
EC/TDS
for current temperature
Screens from DY9812V3 received from Mech.