Hanna Instruments HI9810342 Скачать руководство пользователя страница 9

9

CA

rE

 & 

mA

iNTENANCE

10.  CARE & MAINTENANCE

•  Never immerse the tester over the maximum immersion level.
•  Fresh buffers should be used for each calibration. Once the sachets are opened the buffer value can change over time.
•  If the electrode is slow or sluggish, soak it in cleaning solution for 20 minutes. Rinse with water and hydrate the 

electrode in storage solution for a minimum of 30 minutes before calibrating.

•  If measurements are taken successively, rinse the probe thoroughly in distilled or deionized water to eliminate 

cross-contamination between measurements.

Storage

To ensure a quick response, the glass tip and junction should be kept hydrated.

1.  When not in use, add a few drops of storage solution to storage cap. If storage solution is not available, pH 4.01 or 

pH 7.01 buffer can be used.

Note:

 Do not store the electrode in distilled or deionized water.

2.  Place the probe tip into the cap and gently tighten the collar.

11.  WARNING & ERROR MESSAGES

----

WRNG

“---- WRNG” displayed during calibration. 

Invalid buffer. 

Check the buffer value and use fresh buffer.

12.02

23.7

Measured value displayed blinking.

Measured value is out of electrode range. 

Clean the electrode to improve condition.

!

8.02

23.7

Battery icon (

!

) displayed blinking.

Battery is below 10 %, replace the battery.

BAtt

DEAD

“BAtt DEAD” and tester turns off.

Depleted battery, replace the battery.

12.  ABBREVIATIONS

ATC 

Automatic Temperature Compensation

GLP 

Good Laboratory Practice

MTC 

Manual Temperature Compensation

NIST  National Institute of Standards and Technology

Содержание HI9810342

Страница 1: ...INSTRUCTION MANUAL HI9810342 Wireless pH Tester for Milk with built in specialized electrode Foodcare...

Страница 2: ...a Hanna Instruments product Please read this instruction manual carefully before using this instrument This manual will provide you with the necessary information for correct use of this instrument a...

Страница 3: ...ffer Set 6 pH Resolution 6 Bluetooth Wireless Mode 6 Bluetooth Pairing 7 Date Time 7 7 Bluetooth 7 8 Hanna Lab Application 7 Tag a Measurement 7 9 Calibration 7 Preparation Guidelines 7 Procedure 7 St...

Страница 4: ...pH 0 01 or 0 1 pH mV 0 1 or 1 mV Temperature 0 1 C 0 1 F Accuracy pH 0 05 pH Temperature 0 5 C 0 9 F Calibration Up to three points or four points Automatic buffer recognition with Standard buffers H...

Страница 5: ...342 features a non porous glass body that is easy to clean and disinfect The conical tip allows for easy penetration into semisolids ideal for milk and milk products like yogurt Specialized low temper...

Страница 6: ...d battery type Dispose of old battery in accordance with local regulations 6 SETUP The Setup button is located inside the battery compartment After Setup configuration replace the cover Setup Menu Nav...

Страница 7: ...able mode it will appear in the list of Available Devices Within the application tap Connect to pair the tester with the device All readings are transmitted directly to the application Tag a Measureme...

Страница 8: ...e in the second buffer rinse beaker then in the second calibration buffer Wait until the measurement is stored and the stability indicator disappears 7 After the second point has been stored CAL SAVE...

Страница 9: ...not in use add a few drops of storage solution to storage cap If storage solution is not available pH 4 01 or pH 7 01 buffer can be used Note Do not store the electrode in distilled or deionized water...

Страница 10: ...d for the environment in which it is used Any variation introduced by the user to the supplied equipment may degrade the tester s performance For your and the tester s safety do not use or store it in...

Страница 11: ...12208A 04 This device complies with Industry Canada license exempt RSS standard s Operation is subject to the following two conditions 1 this device may not cause interference and 2 this device must a...

Страница 12: ...World Headquarters Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA www hannainst com MAN9810342 Printed in ROMANIA...

Отзывы: