21
METHOD PROCEDURE
MEASUREMENT PROCEDURE
Note:
If tutorial mode is disabled, follow the measurement procedure below. If the
tutorial mode is enabled, press
Measure
and follow the messages on the screen.
• Fill the cuvette with 10 mL of unreacted sample (up to
the mark).
• Add 5 drops of
HI96779A-0
Chlorine Dioxide Reagent A.
• Replace the plastic stopper and the cap. Shake gently
for 30 seconds.
• Insert the cuvette into the holder and ensure that the
notch on the cap is positioned securely in the groove.
• Press
Zero
. The display will show a 30 second countdown prior to zeroing. To skip
the timer, press
Zero
twice. The meter will display “-0.0-” when the meter is zeroed
and ready for measurement.
10 mL
×
5