20
METHOD PROCEDURE
8. METHOD PROCEDURE
REQUIRED REAGENTS
POWDER
Code Description
Quantity
HI95761-0
Total Chlorine ULR Reagent
1 packet
REAGENT SETS
HI95761-01
Total Chlorine ULR Reagent - 100 tests
HI95761-03
Total Chlorine ULR Reagent - 300 tests
For other accessories see ACCESSORIES section.
MEASUREMENT PROCEDURE
Note:
If tutorial mode is disabled follow the measurement procedure below. If the
tutorial mode is enabled, press
Measure
and follow the messages on the screen.
• Fill the cuvette with 10 mL of unreacted sample (up to the
mark). Replace the plastic stopper and the cap.
• Insert the cuvette into the holder and ensure that the notch
on the cap is positioned securely in the groove.
• Press
Zero
. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
• Remove the cuvette.
• Add one packet of
HI95761-0
Total Chlorine
ULR Reagent. Replace the plastic stopper and
the cap. Shake gently for 20 seconds.
10 mL