7.2. CUVETTE & CYLINDER PREPARATION
Proper mixing is very important for reproducibility of the measurements. The proper
mixing technique is listed in the method procedure.
Invert the cuvette a couple of times or for a specified time: hold the cuvette in the
vertical position. Turn the cuvette upside-down and wait for all of the solution to flow
to the cap end, then return the cuvette to the upright vertical position and wait for all
of the solution to flow to the cuvette bottom. This is one inversion. The correct speed
for this mixing technique is 10-15 complete inversions in 30 seconds. This mixing
technique is indicated with “invert to mix” and the following icon:
In order to avoid reagent leaking and to obtain more accurate measurements,
close the cuvette first with the supplied HDPE plastic stopper
and then the
black cap.
Whenever the cuvette is placed into the measurement holder, it must be dry outside
and free of fingerprints, oil or dirt. Wipe it thoroughly with
HI731318
microfiber
cleaning cloth or a lint-free wipe prior to insertion.