background image

12 

Backlight

Values: 0 to 100 %

Press the 

Modify

 key to access the backlight intensity.

Use the 

 

 keys to increase or decrease the value.

Press the 

Accept

 key to confirm or the 

 key to 

return to the 

Setup

 menu without saving the new 

value.

Date/Time

Press the 

Modify

 key to change the date/time.

Press the 

 

 keys to highlight the value to be 

modified (year, month, day, hour, minute or second).

Press 

Edit

 to modify the highlighted value. Use the 

 

 keys to change the value.

Time Format

Option: AM/PM or 24-hour

 

Press the functional key to select the desired time 

format.

GENERAL OPERA

TIONS

Contrast

Values: 0 to 100 %

Press the 

Modify

 key to change the display’s contrast.

Use the 

 

 keys to increase or decrease the value.

Press the 

Accept

 key to confirm the value or the 

 

key to return to the 

Setup

 menu without saving the 

new value.

6.4. GENERAL SETUP

Press the 

 key to enter the menu. Use the 

 

 keys to select 

Setup

 and press 

Select

. Use the 

 

 keys to highlight desired option.

Press 

No

 or the 

 key to return to the previous screen.

Press 

Yes

 to delete selected log.

Press 

Del All

 to erase all the logged data.

If 

Del All

 is pressed follow the prompt to confirm.

Press 

Yes

 to delete all logged data, 

No

 or the 

 key to return to the log recall.

Press the 

Accept

 key to confirm or the 

 key to 

return to the previous screen

.

Содержание HI97734

Страница 1: ...INSTRUCTION MANUAL HI97734 FreeandTotal ChlorineHighRange Photometer...

Страница 2: ...nstrument as well as a precise idea of its versatility If you need additional technical information do not hesitate to e mail us at tech hannainst com or view our worldwide contact list at www hannain...

Страница 3: ...EAGENTS ACCESSORIES 6 6 CONTEXTUAL HELP 6 7 BATTERY MANAGEMENT 7 PHOTOMETER 7 1 METHOD SELECTION 7 2 COLLECTING AND MEASURING REAGENTS AND SAMPLES 7 2 1 PROPER USE OF POWDER PACKET 7 2 2 PROPER USE OF...

Страница 4: ...te cap 2 pcs Plastic stopper 2 pcs 1 5V AA Alkaline batteries Instruction manual Meter quality certificate Note Save all packing material until you are sure that the instrument works correctly Any dam...

Страница 5: ...f After 15 minutes of inactivity 30minutesbeforeaREADmeasurement Battery type 1 5 V AA Alkaline 3 pcs Battery life 800 measurements without backlight Environment 0 to 50 C 32 to 122 F 0 to 100 RH non...

Страница 6: ...any time and apply a user calibration if necessary Hanna s CAL Check cuvettes are made with NIST traceable standards The built in tutorial mode guides users step by step through the measurement proces...

Страница 7: ...unctiondisplayedabovethemontheLCD Press and hold to power off on Press briefly to return to the previous screen Press to access the menu screen Press to display the context sensitive help menu DESCRIP...

Страница 8: ...extinction coefficient at wavelength c molar concentration of the substance d optical path through the substance Therefore the concentration c can be calculated from the absorbance of the substance a...

Страница 9: ...is higher measurement stability and less wavelength error A focusing lens collects all of the light that exits the cuvette eliminating errors from cuvette imperfections and scratches eliminating the...

Страница 10: ...libration standards is obtained please update the certificate value 4 Insert the A ZERO CAL Check Cuvette A then press Next to continue The Please Wait message will be displayed during the measurement...

Страница 11: ...e of the last user calibration if available or factory calibration To erase the last user calibration and to clear the CAL Check press Clear and follow the prompts Press Yes to erase and return to the...

Страница 12: ...ime format GENERAL OPERATIONS Contrast Values 0 to 100 Press the Modify key to change the display s contrast Use the keys to increase or decrease the value Press the Accept key to confirm the value or...

Страница 13: ...s the functional key to select the desired decimal separator The decimal separator is used on the measurement screen Beeper Option Enable or Disable When enabled a short beep is heard every time a key...

Страница 14: ...battery full battery below 10 replace the batteries soon battery is low replace the batteries with new ones 6 6 CONTEXTUAL HELP The HI97734 offers an interactive contextual help mode that assists the...

Страница 15: ...erve battery the meter will turn off automatically after 15 minutes of inactivity If a zero reading has been done but not a read auto off time is increased to 30 minutes 7 2 2 PROPER USE OF SYRINGE a...

Страница 16: ...ly tapping the cuvette Do not let the reacted sample stand too long after reagent has been added For best accuracy respect the timings described in each specific method It is possible to take multiple...

Страница 17: ...sts HI93734 03 Free and Total Chlorine Reagent HR 300 tests For other accessories see page 25 MEASUREMENT PROCEDURE Select the Free Chlorine HR method using the procedure described in the Method Selec...

Страница 18: ...the cuvette into the holder and ensure that the notch on the cap is positioned securely in the groove FREE CHLORINE HR Insert the cuvette into the holder and ensure that the notch on the cap is positi...

Страница 19: ...CO3 2 sample pH 9 0 In both cases it will not reliably develop the full amount of color or it may rapidly fade negative error To resolve this neutralize the sample with diluted HCl Acidity above 150 m...

Страница 20: ...other accessories see page 25 MEASUREMENT PROCEDURE Select the Total Chlorine HR method using the procedure described in the Method Selection section see page 15 Note If tutorial mode is disabled fol...

Страница 21: ...Shake gently for 20 seconds Insert the cuvette into the holder and ensure that the notch on the cap is positioned securely in the groove Press the Zero key The display will show 0 0 when the meter is...

Страница 22: ...elop the full amount of color or it may rapidly fade negative error To resolve this neutralize the sample with diluted NaOH Hardness in case of water with hardness greater than 500 mg L CaCO3 shake th...

Страница 23: ...not contain any debris Light High There is too much light to perform a measurement Please check the preparation of the zero cuvette 9 ERROR DESCRIPTIONS Ambient temperature out of limits The meter is...

Страница 24: ...instrument off by pressing and holding the key Remove the battery cover by turning it counterclockwise Remove the old batteries replace them with three new 1 5V AA batteries Replace the battery cover...

Страница 25: ...HI7101412 blue carrying case for HI977XX and 2 CAL Check cuvettes HI731318 cloth for wiping cuvettes 4 pcs HI731331 glass cuvettes 4 pcs HI731336N cap for cuvette 4 pcs HI740236 5 mL graduated syringe...

Страница 26: ...cycling Ensuring proper product and battery disposal prevents potential negative consequences for the environment and human health which may be caused by inappropriate handling For more information co...

Страница 27: ...prescribed maintenance is not covered If service is required contact your local Hanna Instruments Office If under warranty report the model number date of purchase serial number engraved on the botto...

Страница 28: ...nts Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA www hannainst com Local Office Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA P...

Отзывы: