20
METHOD PROCEDURE
8. METHOD PROCEDURE
REQUIRED REAGENTS
Code Description Quantity
HI93729-0
Fluoride LR Reagent
4 mL
REAGENT SETS
HI93729-01
Fluoride LR Reagent - 100 tests
HI93729-03
Fluoride LR Reagent - 300 tests
For other accessories see ACCESSORIES section.
MEASUREMENT PROCEDURE
Note:
If tutorial mode is disabled, follow the measurement procedure below. If the
tutorial mode is enabled, press
Measure
and follow the messages on the screen.
• Use a syringe and add 2 mL of
HI93729-0
Fluoride LR Reagent to two cuvettes.
• Use a plastic pipette to fill one cuvette
(#1) with DIW (up to the 10 mL mark).
Replace the plastic stopper and the cap.
Invert several times to mix.
• Use a plastic pipette to fill the second
cuvette (#2) with sample (up to the
10 mL mark). Replace the plastic stopper
and the cap. Invert several times to mix.
• Insert the first cuvette (#1) into the holder and
ensure that the notch on the cap is positioned
securely in the groove.
2 mL
10 mL
# 1
10 mL
# 2
# 1