20
METHOD PROCEDURE
8. METHOD PROCEDURE
REQUIRED REAGENTS
Code Description Quantity
HI93714A-0
Cyanide Reagent A
1 spoon
HI93714B-0
Cyanide Reagent B
1 packet
HI93714C-0
Cyanide Reagent C
1 packet
REAGENT SETS
HI93714-01
Cyanide Reagent - 100 tests
HI93714-03
Cyanide Reagent - 300 tests
For other accessories see ACCESSORIES section.
MEASUREMENT PROCEDURE
Note:
If tutorial mode is disabled, follow the measurement procedure below. If the
tutorial mode is enabled, press
Measure
and follow the messages on the screen.
• Fill the cuvette with 10 mL of unreacted sample (up
to the mark). Replace the plastic stopper and the cap.
• Insert the cuvette into the holder and ensure that the
notch on the cap is positioned securely in the groove.
• Press
Zero
. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
• Remove the cuvette and add 1 level spoon of
HI93714A-0
Cyanide Reagent A. Remember to close
the reagent bottle immediately after use.
10 mL