background image

8

DESCRIPTION

5.2.  FUNCTIONAL DESCRIPTION

1. ON/OFF power button 3. Liquid Crystal Display (LCD)

5. Indexing mark

2. Keypad

4. Cuvette holder

6. Battery cover

Keypad Description

The keypad contains 3 direct keys and 3 functional keys with the following functions:

 Press the functional key to perform the function displayed above it on the LCD.

 Press and hold to power off/on. Press briefly to return to the previous screen.

 Press to access the menu screen.

 Press to display the context-sensitive help menu.

Содержание HI97709

Страница 1: ...INSTRUCTION MANUAL HI97709 Manganese High Range Photometer ...

Страница 2: ...g a Hanna Instruments product Pleasereadthisinstructionmanualcarefullybeforeusingthisinstrument This manual will provide you with the necessary information for correct use of this instrument as well as a precise idea of its versatility If you need additional technical information do not hesitate to e mail us at tech hannainst com or view our worldwide contact list at www hannainst com Dear Custome...

Страница 3: ... 6 1 METER VALIDATION CAL CHECK CALIBRATION 11 6 2 CHEMICAL FORMULA UNIT CONVERSION 12 6 3 GLP 13 6 4 LOGGING DATA LOG RECALL 13 6 5 GENERAL SETUP 14 6 6 REAGENTS ACCESSORIES 16 6 7 CONTEXTUAL HELP 17 6 8 BATTERY MANAGEMENT 17 7 PHOTOMETER 18 7 1 PROPER USE OF POWDER PACKET 18 7 2 CUVETTE PREPARATION 18 8 METHOD PROCEDURE 20 9 WARNING ERROR DESCRIPTIONS 22 10 BATTERY REPLACEMENT 24 11 ACCESSORIES ...

Страница 4: ...te A HI97709B CAL Check Cuvette B for Manganese High Range Cloth for wiping cuvettes Scissors 1 5V AA Alkaline batteries CAL Check standard certificate Instrument quality certificate Instruction manual Each HI97709 is delivered in a cardboard box and is supplied with Sample cuvette 2 pcs Sample cuvette cap 2 pcs Plastic stopper 2 pcs 1 5V AA Alkaline batteries Instrument quality certificate Instru...

Страница 5: ...d rinse with plenty of water If reagent contacts skin rinse the affected area thoroughly with water Avoid breathing released vapors Waste disposal For proper disposal of reagent kits and reacted samples contact a licensed waste disposal provider 3 ABBREVIATIONS mg L milligrams per liter ppm mL milliliter C degree Celsius F degree Fahrenheit EPA US Environmental Protection Agency GLP Good Laborator...

Страница 6: ...er wavelength accuracy 1 0 nm Light detector Silicon photocell Cuvette type Round 24 6 mm diameter 22 mm inside Additional Specifications Auto logging 50 readings Display 128 x 64 pixel B W LCD with backlight Auto off After 15 minutes of inactivity 30 minutes before a READ measurement Battery type 1 5 V AA Alkaline 3 pcs Battery life 800 measurements without backlight Environment 0 to 50 C 32 to 1...

Страница 7: ...rmance of the instrument at any time and apply a user calibration if necessary Hanna Instruments CAL Check cuvettes are made with NIST traceable standards The built in tutorial mode guides users step by step through the measurement process It includes all steps required for sample preparation the required reagents and quantities The HI97709 meter measures the manganese content in water samples in ...

Страница 8: ...over Keypad Description The keypad contains 3 direct keys and 3 functional keys with the following functions Press the functional key to perform the function displayed above it on the LCD Press and hold to power off on Press briefly to return to the previous screen Press to access the menu screen Press to display the context sensitive help menu ...

Страница 9: ...sorbed by atoms molecules or crystal lattices Photometric chemical analysis is based on specific chemical reactions between a sample and reagent to produce a light absorbing compound If pure absorption occurs the fraction of light absorbed depends both on the optical path length through the matter and on the physical chemical characteristics of the substance according to the Lambert Beer Law If al...

Страница 10: ... luminous efficiency providing more light while using less power They also produce little heat which could otherwise affect electronic stability LEDs are available in a wide array of wavelengths whereas tungsten lamps have poor blue violet light output Improved optical filters ensure greater wavelength accuracy and allow a brighter stronger signal to be received The end result is higher measuremen...

Страница 11: ...om direct sunlight by keeping them in the original packing Store between 5 and 30 C 41 to 86 F do not freeze To perform a CAL Check 1 Press CAL Check from measurement mode The Not Available message or the date time and status of the last CAL Check will be displayed on the screen 2 Press Check to start a new CAL Check Press the key at any time to abort the validation process 3 Use the functional ke...

Страница 12: ...screen OUT OF SPECIFICATION A user calibration is not allowed the measured value is outside of the tolerance window Check the certified value expiration date and clean the outside of the cuvette Repeat the CAL Check procedure If this error continues contact your nearest Hanna Instruments Customer Service Center 6 2 CHEMICAL FORMULA UNIT CONVERSION Chemical formula and unit conversion factors are p...

Страница 13: ...ll measurements Every time a measurement is made the data is automatically saved The data log can hold 50 individual measurements When the data log is full 50 data points the meter will rewrite the oldest data point Viewing and deleting the data is possible using the Log Recall menu Press the key to enter the menu Use the functional keys to select Log Recall and press Select Use the functional key...

Страница 14: ...rease or decrease the value Press Accept to confirm or the key to return to the Setup menu without saving the new value Contrast Option 0 to 100 Press Modify to change the display s contrast Use the functional keys to increase or decrease the value Press Accept to confirm the value or the key to return to the Setup menu without saving the new value Date Time Press Modify to change the date and tim...

Страница 15: ...format Press Accept to confirm or the key to return to the Setup menu without saving the new format Decimal Separator Option Comma or Period Press the functional key to select the desired decimal separator The decimal separator is used on the measurement screen Language Press Modify to change the language Use the functional keys to select the desired language Press Accept to choose one of the lang...

Страница 16: ...ded step by step through the measurement procedure Meter Information Press Select to view the model serial number firmware version and selected language Press the key to return to the Setup menu Restore Factory Settings Press Select to reset to factory settings Press Accept to confirm or Cancel to exit without restoring the factory settings 6 6 REAGENTS ACCESSORIES Pressthe keytoenterthemenu Useth...

Страница 17: ... to the previous screen 6 8 BATTERY MANAGEMENT The meter will perform an auto diagnostic test when it is powered on During this test the Hanna Instruments logo will appear on the LCD If the auto diagnostic test was successful the meter is ready for use The battery icon on the LCD will indicate the battery status battery full battery below 10 replace the batteries soon battery is low replace the ba...

Страница 18: ...chnique is listed in the method procedure The mixing method is indicated with shake gently using one of the following icons In order to avoid reagent leaking and to obtain more accurate measurements close the cuvette first with the supplied HDPE plastic stopper and then the black cap Whenever the cuvette is placed into the measurement holder it must be dry outside and free of fingerprints oil or d...

Страница 19: ... in the method It is possible to take multiple readings in a row but it is recommended to take a new zero reading for each sample and to use the same cuvette for zeroing and measurement when possible Discard the sample immediately after the reading has been taken or the glass might become permanently stained All the reaction times reported in this manual are at 25 C 77 F In general the reaction ti...

Страница 20: ...abled press Measure and follow the messages on the screen Fill the cuvette with 10 mL of unreacted sample up to the mark Replace the plastic stopper and the cap Insert the cuvette into the holder and ensure that the notch on the cap is positioned securely in the groove Press Zero The display will show 0 0 when the meter is zeroed and ready for measurement Remove the cuvette Add one packet of HI937...

Страница 21: ...e When the timer ends the meter will perform the reading The instrument displays the results in mg L of manganese Mn Press the key and use the functional keys to select Chemical Form Use the functional keys and press Select to change the displayed chemical formula to mg L potassium permanganate KMnO4 and permanganate MnO4 INTERFERENCES Interference may be caused by Magnesium above 100000 mg L Chlo...

Страница 22: ...d the zero cuvettes are inverted Swap the cuvettes and repeat the measurement There is either too much light or the instrument can not adjust the light level Please check the preparation of the zero cuvette and that the sample does not contain any debris The meter is either overheating or its temperature has dropped too low to operate within published accuracy specifications The meter must be betw...

Страница 23: ...he meter If the issue persists please contact Hanna Instruments technical support Battery level is too low for the meter to function properly Replace the batteries with new ones Tutorial mode has been enabled in the Setup menu Press Continue and follow the prompt on the screen Tutorial mode can be disabled in the Setup menu A critical error has occurred Restart the meter If the issue persists plea...

Страница 24: ...ent s batteries follow these steps Turn the instrument off by pressing and holding the key Remove the battery cover by turning it counterclockwise Remove the old batteries replace them with three new 1 5V AA batteries Replace the battery cover turn it clockwise to close ...

Страница 25: ...eagent 300 tests 11 2 OTHER ACCESSORIES Code Description HI7101412 blue carrying case for HI977xx and 2 CAL Check cuvettes HI731318 cloth for wiping cuvettes 4 pcs HI731331 glass cuvette 4 pcs HI731336N cap for glass cuvette 4 pcs HI93703 50 cuvette cleaning solution 230 mL HI97709 11 CAL Check standards for Manganese HR cuvette kit ...

Страница 26: ...t in which it is used Any variation introduced by the user to the supplied equipment may degrade the meter s performance For yours and the meter s safety do not use or store the meter in hazardous environments WARRANTY The HI97709 is warranted for two years against defects in workmanship and materials when used for its intended purpose and maintained according to instructions This warranty is limi...

Страница 27: ...Hanna Instruments reserves the right to modify the design construction or appearance of its products without advance notice ...

Страница 28: ...World Headquarters Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA www hannainst com MAN97709 Printed in ROMANIA ...

Отзывы: