METHOD PROCEDURE
8. METHOD PROCEDURE
REQUIRED REAGENTS
Code Description Quantity
HI93706A‑0
Phosphorus Reagent A
10 drops
HI93706B‑0
Phosphorus Reagent B
1 packet
REAGENT SETS
HI93706‑01
Phosphorus Reagent ‑ 100 tests
HI93706‑03
Phosphorus Reagent ‑ 300 tests
For other accessories see ACCESSORIES section.
MEASUREMENT PROCEDURE
Note:
If tutorial mode is disabled follow the measurement procedure below. If the
tutorial mode is enabled, press
Measure
and follow the messages on the screen.
• Fill the cuvette with 10 mL of unreacted sample (up
to the mark). Replace the plastic stopper and the cap.
• Insert the cuvette into the holder and ensure that the
notch on the cap is positioned securely in the groove.
• Press
Zero
. The display will show “‑0.0‑” when the meter is zeroed and ready for
measurement.
• Remove the cuvette.
• Add 10 drops of
HI93706A‑0
Phosphorus
Reagent A.
• Add one packet of
HI93706B‑0
Phosphorus
Reagent B to the cuvette. Replace the plastic
stopper and the cap. Shake gently until
completely dissolved.
10 mL
×
10