17
METHOD PROCEDURE
10 mL
8. METHOD PROCEDURE
REQUIRED REAGENTS
Code
Description Quantity
HI93700A‑0
Ammonia Low Range Reagent A
4 drops
HI93700B‑0
Ammonia Low Range Reagent B
4 drops
REAGENT SETS
HI93700‑01
Ammonia LR Reagent ‑ 100 tests
HI93700‑03
Ammonia LR Reagent ‑ 300 tests
For other accessories see page 22.
MEASUREMENT PROCEDURE
Note
: If tutorial mode is disabled follow the measurement procedure below. If the
tutorial mode is enabled, press
Measure
and follow the messages on the screen.
• Fill the cuvette with 10 mL of unreacted sample (up to
the mark). Replace the plastic stopper and the cap.
• Place the cuvette into the holder and ensure that the
notch on the cap is positioned securely into the groove.
• Press the
Zero
key. The display will show “
-0.0-
” when the meter is zeroed and
ready for measurement.
• Remove the cuvette.