43
METHOD PROCEDURE
8.9. IODINE
REQUIRED REAGENTS
Code Description Quantity
HI93718-0
Iodine Reagent
1 packet
REAGENT SETS
HI93718-01
Iodine Reagent - 100 tests
HI93718-03
Iodine Reagent - 300 tests
For other accessories see ACCESSORIES section.
MEASUREMENT PROCEDURE
• Select the
Iodine
method using the procedure described in the METHOD SELECTION
section.
Note:
If tutorial mode is disabled, follow the measurement procedure below. If the
tutorial mode is enabled, press
Measure
and follow the messages on the screen.
• Fill a cuvette with 10 mL of unreacted sample (up
to the mark). Replace the plastic stopper and the
cap.
• Insert the cuvette into the holder and ensure that
the notch on the cap is positioned securely in the
groove.
• Press
Zero
. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
10 mL