Hanna Instruments HI96786 Скачать руководство пользователя страница 10

11•

Then the date of last calibration (e.g.:

“01.08.2009”

) appears on the display, or

“01.01.2009”

 if the factory calibration

was selected before. In both cases the year
number is blinking, ready for date input.

12•

Press 

GLP/

 to edit the desired year

(2009-2099). If the key is kept pressed,
the year number is automatically increased.

13•

When the correct year has been set, press

ZERO/CFM

 or 

READ/ /TIMER

 to confirm.

Now the display will show the month
blinking.

14•

Press 

GLP/

 to edit the desired month

(01-12). If the key is kept pressed, the
month number is automatically increased.

15•

When the correct month has been set,
press 

ZERO/CFM

 or 

READ/ /TIMER

 to

confirm. Now the display will show the
day blinking.

16•

Press 

GLP/

 to edit the desired day

(01-31). If the key is kept pressed, the
day number is automatically increased.

Note:

It is possible to change the editing from
day to year and to month by pressing

READ/ /TIMER.

17•

Press 

ZERO/CFM

 

to save the calibration

date.

18•

The instrument displays “

Stor

” for one

second and the calibration is saved.

19•

The instrument will return automatically
to 

measurement mode

 by displaying dashes

on the LCD.

R

EAGENT SETS

HI 93728-01

Reagents for 100 tests

HI 93728-03

Reagents for 300 tests

OTHER ACCESSORIES

HI 96786-11

CAL CHECK™ Standard Cuvettes (1 set)

HI 721310

9V battery (10 pcs)

HI 731318

Cloth for wiping cuvettes (4 pcs)

HI 731331

Glass cuvettes (4 pcs)

HI 731335

Caps for cuvettes (4 pcs)

HI 93703-50

Cuvette cleaning solution (230 mL).

Warranty

HI 96786

 is warranted for two years against defects in workmanship and

materials when used for its intended purpose and maintained according
to the instructions.
This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed
maintenance are not covered.
If service is required, contact your dealer. If under warranty, report the
model number, date of purchase, serial number and the nature of the
failure. If the repair is not covered by the warranty, you will be notified
of the charges incurred.
If the instrument is to be returned to Hanna Instruments, first obtain a
Returned Goods Authorization Number from the Customer Service Department
and then send it with shipment costs prepaid. When shipping any
instrument, make sure it is properly packaged for complete protection.
To validate your warranty, fill out and return the enclosed warranty card
within 14 days from the date of purchase.

Recommendations for Users

Before using these products, make sure that they are entirely suitable for your specific application and for
the environment in which they are used.
Operation of these instruments may cause  unacceptable interferences to other electronic equipments, this
requiring the operator to take all necessary steps to correct interferences.
Any variation introduced by the user to the supplied equipment may degrade the instrument’s EMC
performance.
To avoid damages or burns, do not put the instrument in microwave oven. For yours and the instrument
safety do not use or store the instrument in  hazardous environments.

Hanna Instruments 

reserves the right to modify the

design, construction and appearance of its products

without advance notice.

For additional information, contact your

dealer or the nearest

Hanna Customer Service Center.

To find the Hanna Office in your area,

visit our web site

w w w . h a n n a i n s t . c o m

IST96786   12/10

Accessories:

3

Warning:

 do not validate or calibrate the instrument with standard

solutions other than the Hanna 

CAL CHECK™

 Standards, otherwise

erroneous results will be obtained.
For accurate validation and calibration results, please perform tests at
room temperature (18 to 25°C; 64.5 to 77.0°F).

Use the Hanna CAL CHECK™ cuvettes (see
“Accessories”) to validate or calibrate instruments.

Validation and Calibration procedures

VALIDATION

1•

Turn the meter on by pressing 

ON/OFF

.

2•

When the beeper sounds briefly and the
LCD displays dashes, the meter is ready.

3•

Place the 

CAL CHECK™

 Standard

HI 96786-11

 Cuvette A into the holder

and ensure that the notch on the cap
is positioned securely into the groove.

4•

Press 

ZERO/CFM

 and the lamp, cuvette

and detector icons will appear on the display,
depending on the measurement phase.

5•

After a few seconds the display will show

“-0.0-”

. The meter is now zeroed and

ready for validation.

6•

Remove the cuvette.

7•

Place the 

CAL CHECK™

 Standard

HI 96786-11

 Cuvette B into the holder

and ensure that the notch on the cap is
positioned securely into the groove.

8•

Press 

CAL CHECK

 key and the lamp,

cuvette and detector icons together with

“CAL CHECK”

 will appear on the display,

depending on the measurement phase.

Validation 

4-5

7

8

9

Calibration 

9-10

11-13

 or

14-15

 or

16

17

LAST CALIBRATION DATE

1•

Press 

GLP/

 to enter 

GLP mode

. The

calibration month and day will appear on
the main display and the year on the
secondary display.

2•

If no calibration was performed, the factory
calibration message, 

“F.CAL”

 will appear

on the main display and the instrument
returns to 

measurement mode

 after three

seconds.

GLP

Last

Calibration

Date

 

In 

GLP mode

, the last calibration date can be verified and the factory

calibration can be restored.

Factory

Calibration

Restore

 

3-4

2

3-4

Battery management

To save the battery, the instrument shuts down after 10 minutes of non-use
in 

measurement mode 

and after 1 hour of non-use in 

calibration mode.

If a valid measurement was displayed before
auto-shut off, the value is displayed when the
instrument is switched on. The blinking 

“ZERO”

means that a new zero has to be performed.
One fresh battery lasts for arround 750 measurements, depending on the
light level.
The remaining battery capacity is evaluated at the instrument startup
and after each measurement.
The instrument displays a battery indicator with three levels as follows:
• 3 lines for 100 % capacity
• 2 lines for 66 % capacity
• 1 line for 33 % capacity
• Battery icon blinking if the capacity is under 10 %.
If the battery is empty and accurate measurements can’t be taken any
more, the instrument shows 

“dEAd bAtt”

 and turns off.

To restart the instrument, the battery must be replaced with a fresh one.
To replace the instrument’s battery, follow the steps:
• Turn the instrument off by pressing 

ON/OFF

.

• Turn the instrument upside down and remove the battery cover by

turning it counterclockwise.

• Extract the battery from its location and replace it with a fresh one.
• Insert back the battery cover and turn it clockwise to close.

9•

At the end of the measurement the display
will show the validation standard value.
The reading should be within specifications as
reported on the 

CAL CHECK™

 Standard

Certificate. If the value is found out of
specifications, please check that the cuvettes
are free of fingerprints, oil or dirt and repeat
validation. If results are still found out of
specifications then recalibrate the instrument.

CALIBRATION

Note:

It is possible to interrupt the calibration
procedure at any time by pressing 

CAL

CHECK

 or 

ON/OFF

 keys.

1•

Turn the meter on by pressing 

ON/OFF

.

2•

When the beeper sounds briefly and the
LCD displays dashes, the meter is ready.

3•

Press and hold 

CAL CHECK 

for three seconds

to enter 

calibration mode.

 The display will

show 

“CAL”

 during calibration procedure. The

blinking 

“ZERO”

 asks for instrument zeroing.

4•

Place the 

CAL CHECK™

 Standard

HI 96786-11

 Cuvette A into the cuvette

holder and ensure that the notch on the cap
is positioned securely into the groove.

5•

Press 

ZERO/CFM

 and the lamp, cuvette

and detector icons will appear on the display,
depending on the measurement phase.

6•

After a few seconds the display will show

“-0.0-”

. The meter is now zeroed and

ready for calibration. The blinking 

“READ”

asks for reading calibration standard.

7•

Remove the cuvette.

8•

Place the 

CAL CHECK™

 Standard

HI 96786-11

 Cuvette B into the holder

and ensure that the notch on the cap is
positioned securely into the groove.

9•

Press 

READ/ /TIMER

 and the lamp,

cuvette and detector icons will appear on
the display, depending on the measurement
phase.

10•

The instrument will show for three seconds
the 

CAL CHECK™

 standard value.

Note:

If the display shows 

“STD HIGH”

, the

standard value was too high. If the
display shows 

“STD LOW”

, the standard

value was too low. Verify that both 

CAL

CHECK™

 Standard 

HI 96786-11

 Cu-

vettes, A and B are free from fingerprints
or dirt and that they are inserted cor-
rectly.

11•

At the end of measurement, the instrument
directly displays concentration in mg/L of
nitrate on the LCD. To convert the reading
to mg/L of nitrate-nitrogen, multiply by
the factor 0.226.

INTERFERENCES:

Interference may be caused by:
Ammonia and amines, as urea and primary
aliphatic amines. Chloride above 100 mg/
L (negative interference). Chlorine above 2
mg/L (positive interference). Copper (it must
be absent). Iron (III) (positive interference). Strong oxidizing and
reducing substances. Sulfide (it must be absent).

Note

: To ensure accurate results, preform the tests at room

temperature, between 18 °C and 28 °C (65 °F to 83 °F).

For best result

: Intensely colored samples or suspended matter in

large amounts will cause interference. Therefore sample should be
adequately treated before performing the test.

FACTORY CALIBRATION RESTORE

It is possible to delete the calibration and restore
factory calibration.

1•

Press 

GLP/

 to enter 

GLP mode

.

2•

Press 

READ/ /TIMER

 to enter in the

factory calibration restore screen. The
instrument asks for confirmation of user
calibration delete.

3•

Press 

ZERO/CFM

 to restore the factory

calibration or press 

GLP/

 again to

abort factory calibration restore.

4•

The instrument briefly indicates 

“donE”

upon restoration of factory calibration prior
to returning to 

measurement mode

.

2

1

18

 or

11

3

4

5-6

8

Содержание HI96786

Страница 1: ...INSTRUCTION MANUAL HI 96786 HI 96786 HI 96786 HI 96786 HI 96786 www hannainst com Nitrate ISM ...

Страница 2: ...for the correct use of the instrument Please read it carefully before using the meter If you need additional technical information do not hesitate to e mail us at tech hannainst com Technical specifications Range 0 to 100 mg L Resolution 1 mg L Accuracy 5 mg L 5 of reading 25 C Typical EMC Dev 1 mg L Light Source Tungsten Lamp Light Detector Silicon Photocell with narrow band interference filter 5...

Страница 3: ...nt In GLP mode press to view the next screen 5 ON OFF key to turn the meter on and off 6 Liquid Crystal Display LCD 7 Cuvette alignment indicator 8 Cuvette holder DISPLAY ELEMENTS DESCRIPTION Functional description 1 The measuring scheme lamp cuvette detector appears during different phases of zero or reading measurement 2 Error messages and warnings 3 The battery icon indicates the charge state o...

Страница 4: ...er range A blinking 0 0 indicates that the sample absorbs less light than the zero reference Check the procedure and make sure you use the same cuvette for reference zero and measurement Over Range A flashing value of the maximum concentration indicates an over range condition The concentration of the sample is beyond the programmed range dilute the sample and re run the test Standard Low The stan...

Страница 5: ...ng the cuvette gently and slowly for 50 seconds while taking care not to induce air bubbles A deposit could remain but it dose not affect the measurement Time and way of shaking could sensitively affect the measurement 9 Replace the cuvette into the holder and ensure that the notch on the cap is positioned securely into the groove 10 Press and hold READ TIMER for three seconds and the display will...

Страница 6: ...e cuvette 7 Place the CAL CHECK Standard HI 96786 11 Cuvette B into the holder and ensure that the notch on the cap is positioned securely into the groove 8 Press CAL CHECK key and the lamp cuvette and detector icons together with CAL CHECK will appear on the display depending on the measurement phase Validation 4 5 7 8 11 At the end of measurement the instrument directly displays concentration in...

Страница 7: ...CK Standard HI 96786 11 Cuvette A into the cuvette holder and ensure that the notch on the cap is positioned securely into the groove 5 Press ZERO CFM and the lamp cuvette and detector icons will appear on the display depending on the measurement phase 6 After a few seconds the display will show 0 0 The meter is now zeroed and ready for calibration The blinking READ asks for reading calibration st...

Страница 8: ...P to edit the desired day 01 31 If the key is kept pressed the day number is automatically increased Note It is possible to change the editing from day to year and to month by pressing READ TIMER 17 Press ZERO CFM to save the calibration date 18 The instrument displays Stor for one second and the calibration is saved 19 The instrument will return automatically to measurement modeby displaying dash...

Страница 9: ...nd accurate measurements can t be taken any more the instrument shows dEAd bAtt and turns off To restart the instrument the battery must be replaced with a fresh one To replace the instrument s battery follow the steps Turn the instrument off by pressing ON OFF Turn the instrument upside down and remove the battery cover by turning it counterclockwise Extract the battery from its location and repl...

Страница 10: ...vice Department and then send it with shipment costs prepaid When shipping any instrument make sure it is properly packaged for complete protection To validate your warranty fill out and return the enclosed warranty card within 14 days from the date of purchase Recommendations for Users Before using these products make sure that they are entirely suitable for your specific application and for the ...

Отзывы: