Hanna Instruments HI96751 Скачать руководство пользователя страница 22

22 

or

Note: If the display shows “STD HIGH”, the standard value was too high. If the display 

shows “STD LOW”, the standard value was too low. Verify that both CAL Check

TM

 Standards 

HI96751-11 Cuvettes, A and B are free of fingerprints or dirt and that they are inserted 

correctly.

• Then the date of the last calibration (e.g.: ”01.08.2008”) appears on the display, or 

“01.01.2008” if the factory calibration was selected before. In both cases the year number is 
blinking, ready for date input.

CALIBRA

TION PROCEDURE

•  After measurement the instrument will show for three seconds the 

CAL Check

TM

 Standard value.

DATE INPUT

• Press 

GLP/    

to edit the desired year (2000‑2099). If the key is kept pressed, the year number 

is automatically increased. 

•  When the correct year has been set, press 

ZERO/CFM

 or 

READ / TIMER

 to confirm. Now the 

display will show the month blinking.

Содержание HI96751

Страница 1: ...INSTRUCTION MANUAL HI96751C Sulfate ISM...

Страница 2: ...anna Instruments product Please read this instruction manual carefully before using the instrument This manual will provide you with the necessary information for correct use of the instrument as well...

Страница 3: ...ISION AND ACCURACY PRINCIPLE OF OPERATION FUNCTIONAL DESCRIPTION ERRORS AND WARNINGS GENERAL TIPS FOR AN ACCURATE MEASUREMENT STARTUP MEASUREMENT PROCEDURE VALIDATION PROCEDURE CALIBRATION PROCEDURE G...

Страница 4: ...Meter is supplied complete with Sample Cuvettes and Caps 2 pcs CAL Check standard cuvettes 9V Battery Scissors Cloth for wiping cuvettes Instrument quality certificate Instruction Manual and Quick Ref...

Страница 5: ...by a transparent cup This makes the instruments fulfill field applications Display messages aid the user in routine operation The meter has an auto shut off feature that will turn off the instrument a...

Страница 6: ...with barium chloride crystals Light absorbance of the suspension is measured Environment 0 to 50 C 32 to 122 F max 95 RH non condensing Battery Type 9V 1 pc Auto Shut off After 10 of non use in measu...

Страница 7: ...ght absorbed depends both on the optical path length through the matter and on the physical chemical characteristics of the substance according to the Lambert Beer Law log I Io c d or A c d Where log...

Страница 8: ...The cuvette has a very important role because it is an optical element and thus requires particular attention It is important that both the measurement and the calibration zeroing cuvettes are optica...

Страница 9: ...ser calibration date or restore factory calibration KEYPAD DESCRIPTION ON OFF to turn the meter on and off ZERO CFM to zero the meter prior to measurement to confirm edited values or to confirm factor...

Страница 10: ...nd date icons appear when a date is displayed 8 Four digit main display 9 Measuring units 10 Four digit secondary display FUNCTIONAL DESCRIPTION The instrument shows clear messages when erroneous cond...

Страница 11: ...ng 0 indicates that the sample absorbs less light than the zero reference Check the procedure and make sure you use the same cuvette for reference zero and measurement Over Range A flashing value of t...

Страница 12: ...amp The instrument waits for the lamp to cool down d other errors and warnings Battery low The battery must be replaced soon Dead battery This indicates that the battery is dead and must be replaced O...

Страница 13: ...tom of this concavity must be at the same level of the 10 mL mark Proper use of the powder reagent packet a use scissors to open the powder packet b push the edges of the packet to form a spout c pour...

Страница 14: ...stained All the reaction times reported in this manual are at 25 C 77 F In general the reaction time should be increased dor temperatures lower than 20 C 68 F and decreased for temperatures higher tha...

Страница 15: ...fter the reagents are added the reacted sample is measured When the beeper sounds briefly and the LCD displays dashes the meter is ready The blinking ZERO indicates that the instrument needs to be zer...

Страница 16: ...and detector icons will appear on the display depending on the measurement phase Add the content of one packet of HI93751 0 reagent replace the cap and shake gently the cuvette for 1 minute Remove the...

Страница 17: ...INTERFERENCES Color or suspended matter in large amounts will interfere suspended matter should be removed by previous filtration Organic matter in large amounts may impede the precipitation of barium...

Страница 18: ...e holder and ensure that the notch on the cap is positioned securely into the groove When the beeper sounds briefly and the LCD displays dashes the meter is ready The blinking ZERO asks for instrument...

Страница 19: ...e At the end of the measurement the display will show the validation standard value The reading should be within specifications as reported in the CAL CheckTM Standard Certificate If the value is foun...

Страница 20: ...calibration procedure The blinking ZERO asks for instrument zeroing Place the CAL CheckTM Standard HI96751 11 Cuvette A into the cuvette holder and ensure that the notch on the cap is positioned secur...

Страница 21: ...kTM Standard HI96751 11 Cuvette B into the cuvette holder and ensure that the notch on the cap is positioned securely into the groove Press READ TIMER and the lamp cuvette and detector icons will appe...

Страница 22: ...08 2008 appears on the display or 01 01 2008 if the factory calibration was selected before In both cases the year number is blinking ready for date input CALIBRATION PROCEDURE After measurement the i...

Страница 23: ...ly increased When the correct month has been set press ZERO CFM or READ TIMER to confirm Now the display will show the day blinking Press ZERO CFM to save the calibration date Note It is possible to c...

Страница 24: ...display If no calibration was performed the factory calibration message F CAL will appear on the main display and the instrument returns to measurement mode after three seconds LAST CALIBRATION DATE P...

Страница 25: ...en the instrument is switched on The blinking ZERO means that a new zero has to be performed One fresh battery lasts for around 750 measurements depending on the light level The remaining battery capa...

Страница 26: ...wise to close To replace the instrument s battery follow the steps BATTERY REPLACEMENT ACCESSORIES Reagent Set HI93751 01 Reagents for 100 tests HI93751 03 Reagents for 300 tests Other Accessories HI9...

Страница 27: ...s first obtain a Returned Goods Authorization Number from the Customer Service Depart ment and then send it with shipment costs prepaid When shipping any instrument make sure it is properly packaged f...

Страница 28: ...a Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA www hannainst com Local Office Hanna Instruments USA 270 George Washington Highway Smithfield RI 02917 Phone 800...