27
GLP
• Press
READ /TIMER
to enter in the factory calibration restore screen. The instrument asks for
confirmation of user calibration delete.
• Press
ZERO/CFM
to restore the factory calibration or press
GLP
again to abort factory
calibration restore.
• The instrument briefly indicates
“done”
upon restoration of factory calibration prior to returning
to measurement mode.