12
The instructions listed below should be carefully followed during testing to ensure best accuracy.
• Color or suspended matter in large amounts may cause interference, therefore these should be
removed by treatment with active carbon and by prior filtration.
• For a correct filling of the cuvette: the liquid in the cuvette forms
a concavity on the top; the bottom of this concavity must be at
the same level as the 10 mL mark.
• Proper use of the powder reagent packet:
(a) use scissors to open the powder packet;
(b) push the edges of the packet to form a spout;
(c) pour out the content of the packet.
• It is important that the sample does not contain any debris. This would corrupt the reading.
• Each time the cuvette is used, the cap must be tightened to the same degree.
• Whenever the cuvette is placed into the measurement cell, it must be dry outside, and completely
free of fingerprints, oil or dirt. Wipe it thoroughly with
HI731318
or a lint‑free cloth prior to
insertion.
• Shaking the cuvette can generate bubbles in the sample, causing higher readings. To obtain
accurate measurements, remove such bubbles by swirling or by gently tapping the cuvette.
• Do not let the reacted sample stand too long after reagent is added, or accuracy will be lost.
• It is possible to take multiple readings in a row, but it is recommended to take a new zero
reading for each sample and to use the same cuvette for zeroing and measurement.
GENERAL TIPS FOR AN ACCURA
TE MEASUREMENT
HI93713-0
HI93713-0
HI93713-
0