To compensate the meter for the sample turbidity or color, the measurement takes place in two
phases. First, the meter is zeroed using the unreacted sample. After the reagents are added the
reacted sample is measured.
• Turn the meter on by pressing
ON/OFF
. The display briefly shows all tags on.
MEASUREMENT PROCEDURE
• When the beeper sounds briefly and the LCD displays dashes, the meter is ready. The blinking
“ZERO”
indicates that the instrument needs to be zeroed first.
• Fill one cuvette with 10 mL of unreacted sample, up to the mark.
10mL
10ml
• Add 6 drops of
HI93705A-0
Molybdate reagent. Replace the cap and swirl the solution.
6
93705A
• Wait for 4 minutes, add the content of one packet of
HI93705B-0
Citric acid reagent and shake
until it is completely dissolved.