background image

6

7

INSTRUMENT  PREPARATION

Each meter is supplied complete with 4 each

1.5V AA batteries. Remove the back cover, un-

wrap the batteries and install them while paying

attention to their polarity.
Connect the RH probe to the 7-pin DIN socket

on the top of the meter and fasten the screw

tightly.

Note

: HI 9065 has to be used with just one

probe (RH or temperature) at a time.

HUMIDITY  MEASUREMENTS

Turn the meter on by pressing the ON/OFF key.
Press the RH key to display the RH % measurement. The symbol "H"

is displayed when measuring Relative Humidity (not for  HI9160).

Using these hygrometers is simple. However, the recommendations

below should always be followed.
• The end of the humidity detector should be exposed to a current of

air moving at 0.5 m (20") per second or more.

• In the absence of air movement, the response can be accelerated

by moving the probe.

••

•• The probe sensor must never come into contact with water or other

liquids.
If this should happen, or if condensation causes drops to form on
the surface of the humidity sensor, turn off the instrument and

wait until they have evaporated completely.
In order to accelerate the evaporation process, the humidity sensor

could be exposed to a current of air.

ON/OFF

RH

H

OPERATIONAL  GUIDE

OPERATIONAL  GUIDE

OPERATIONAL  GUIDE

OPERATIONAL  GUIDE

OPERATIONAL  GUIDE

HI9064 & HI9160

HI9065

Range

RH

5.0 to 95.0% RH     5.0 to 95.0% RH

°C

--              0.0 to + 60.0 °C

°F

--              32 to +140 °F

Resolution

RH

0.1% RH

        0.1% RH

°C

--

            0.1 °C

°F

--

           1 °F

Accuracy

RH

±2%

       ±2%

(@25°C/77°F) °C

--

          ±0.4 °C

°F

--

           ±1 °F

for one year, excluding probe error

Typical EMC

±3% RH

       ±3% RH

Deviation

               ±2 °C / ± 4°F

Calibration

RH

Through 2 trimmers on RH probe

Probe

HI70608/2

HI70605/2

Operating Conditions

0 to 50 °C (32 to 122°F)

 

100% RH

Dimensions

Instrument

196x80x60 mm (7.7x3.1x2.4")

RH Probe

L:165 mm D:25 mm with 2 m cable

(L:6.5" D:1" 7' cable)

Battery

4 x 1.5V AA / 500 hours of continuous use

Weight

Instrument

425 g (15 oz.)

Probe

200 g (7 oz.)

SPECIFICATIONS

SPECIFICATIONS

SPECIFICATIONS

SPECIFICATIONS

SPECIFICATIONS

Содержание HI9064

Страница 1: ...HI9065 HI9160 Portable Water Resistant Hygrometers HI 9065 Thermo Hygrometer ON OFF HOLD RH C F RH C F 5 95 0 60 C 32 140 F s H h t t p w w w h a n n a i n s t c o m These Instruments are in Complian...

Страница 2: ...ler Note Save all packing material until you are sure that the instrument functions correctly All defective items must be returned to us in the original packing with the supplied accessories The HI906...

Страница 3: ...064 H 9064 H 9064 H 9064 H 9064 HI I I I I9065 9065 9065 9065 9065 HI 9065 Thermo Hygrometer HI 9064 Hygrometer ON OFF HOLD RH ON OFF C F HOLD RH C F RH 5 95 5 95 0 60 C 32 140 F 1 7 8 9 3 2 5 4 2 3 6...

Страница 4: ...come into contact with water or other liquids If this should happen or if condensation causes drops to form on the surface of the humidity sensor turn off the instrument and wait until they have evapo...

Страница 5: ...he walls of the glass container Li Cl 11 1 Na Cl 75 4 CALIBRATION KIT H 2 O 26 cc L i C l CALIBRATION CALIBRATION CALIBRATION CALIBRATION CALIBRATION C C F TEMPERATURE MEASUREMENTS FOR HI9065 ONLY Con...

Страница 6: ...t in use as the LiCl solution is extremely hygroscopic and tends to capture the humidity present in the air causing the solution to expand in volume and overflow from the container Pour approximately...

Страница 7: ...9065 ONLY Interchangeable precalibrated temperature probes HI765AN Air Temperature Probe HI765FN Surface Temperature Probe HI765LN Liquid Temperature Probe HI765PN Penetration Temperature Probe HI9064...

Страница 8: ...WARRANTY WARRANTY WARRANTY WARRANTY WARRANTY All Hanna Instruments meters are guaranteed for two years against defects in workmanship and materials when used for their intended purpose and maintained...

Страница 9: ...ITERATURE HANNA LITERATURE HANNA LITERATURE LAB RECORDING WATER ANALYSIS HANDBOOK PRINTED IN ITALY ENVIROCARE GENERAL CATALOG These and many others catalogs handbooks and leaflets are available from H...

Отзывы: