7
instrument's power supply is disconnected before the meter is switched
off. Press the RESET button and restart the entire operation.
Power connection
Plug the 12VDC adapter (
HI710005
or
HI710006
) into the DC socket
(#6).
Note:
Make sure the main line is protected by a fuse.
Note: HI8417
uses a rechargeable battery to retain the pH calibra-
tion data. It lasts approximately one month when fully charged. The
battery is automatically charged any time the meter is plugged-in
(either turned on or off).
SPECIFICATIONS
HI8417
Range
pH
0.00 to 14.00
ISE
0.0 to ±399.9mV
ORP
±1999mV
°C
-10.0 to 125.0
Resolution
pH
0.01
ISE
0.1mV
ORP
0.1/1mV
°C
0.1
Accuracy
pH
±0.01
(@ 20°C / 68°F)
ISE
±0.2 mV
ORP
±1 mV
°C
±0.5
Typical EMC
pH
±0.01
Deviation
ISE/ORP
±0.2mV
°C
±0.7
Recorder
pH
0 to 2 V
Output
mV
0 to 2 V
Max. Current
20mA
Offset
±1 pH
Slope
from 85% to 105%
Calibration
Automatic 2 points with 5 memorized buffers
(pH 4.01, 6.86, 7.01, 9.18, 10.01)
Temperature
Manual from 0 to 100°C (32 to 212°F)
Compensation
Automatic from -10 to 125°C (14 to 257°F)
Electrode
HI1131B
glass body combination refillable
pH electrode (
included
)
Temp. Probe
HI7669/2W
Temperature Probe (
included
)
Input Impedance
10
12
ohm
Power
Power socket for 12VDC
Environment
0 to 50°C (32 to 122°F);
95% RH (not condensing)
Dimensions
230Lx170Wx70H mm (9.1x6.7x2.7")
Shipping Weight
1.5 Kg (3.3 lb.)