
69
OXYGEN
, DISSOL
VED
8.16. OXYGEN, DISSOLVED
SPECIFICATIONS
Range
0.0 to 10.0 mg/L (as O
2
)
Resolution
0.1 mg/L
Accuracy
±0.4 mg/L ±3% of reading at 25 °C
Light Source
LED with narrow band interference filter @ 420 nm
Method
Adaptation of the Standard Methods for the Examination of Water and
Wastewater, 18th edition, Azide modified Winkler method.
REQUIRED REAGENTS
Code
Description Quantity
HI93732A‑0
Dissolved Oxygen Reagent A
5 drops
HI93732B‑0
Dissolved Oxygen Reagent B
5 drops
HI93732C‑0
Dissolved Oxygen Reagent C
10 drops
REAGENT SET
HI93732‑01
Reagents for 100 tests
HI93732‑03
Reagents for 300 tests
For other accessories see page 87.
MEASUREMENT PROCEDURE
• Select the
Oxygen (dissolved)
method using the procedure
described in the Method Selection section (see page 17).
• Fill one 60 mL glass bottle completely with the unreacted
sample.
• Replace the cap and ensure that a small part of the sample
spills over.
• Remove the cap and add 5 drops of
HI93732A‑0
and 5 drops of
HI93732B‑0
.
• Add more sample, to fill the bottle completely. Replace the cap
and ensure that a part of the sample spills over.
Note: This ensures no air bubbles have been trapped inside
the bottle. Trapped air bubbles could alter readings.
• Invert the bottle several times until the sample turns orange‑
yellow and a flocculating agent appears.