PROBE MAINTENANCE

General maintenance

After use, rinse the probe with tap water 
and dry it. The pH electrode bulb must be 
kept moist, while the D.O. and EC sensors 
should be stored dry.

 

Maintenance of D.O. sensor

For a top performance probe, replace
the membrane every 2 months and the 
electrolyte once a month. Proceed as
follows:
•  Unscrew the membrane by

turning it counterclockwise.

•  Rinse the spare membrane

with some electrolyte while
shaking it gently. Refill with
clean electrolyte.

•  Gently tap the membrane over a surface

 

to 

ensure that no air bubbles remain

 

trapped.

•  With the sensor facing down, screw the

 

cap clockwise to the end of the threads.

 

Some electrolyte will overflow.

If any deposit scales the sensors, gently brush
the sensor surface with the brush supplied in 
the maintenance kit, while paying attention 
to not damage the plastic body. 

Maintenance of EC sensor

After every series of measurements, rinse
the probe with tap water.
If a more thorough cleaning is required, 
clean the probe with the brush supplied 
in the maintenance kit or a nonabrasive 
detergent. 
After cleaning the probe, always re-
calibrate the instrument. 

Maintenance of pH & pH/ORP sensors

• Remove the protective cap. Do not be

alarmed if any salt deposits are present.
This is normal with electrodes and they
will disappear when rinsed with water.

• Shake the electrode down as you would

do with a clinical thermometer to elimi-
nate any air bubbles inside the glass
bulb.

• If the bulb and/or junction are dry, soak

the electrode in 

HI70300

 storage solu-

tion for at least one hour.

• To minimize clogging and ensure a quick 

response time, the glass bulb and the
junction should be kept moist and not
allowed to dry.

• Replace the solution in the protective cap 

with a few drops of 

HI70300

 storage

solution. Tap water may also be used
for a very short period (couple of days).

NEVER STORE THE ELECTRODE 
IN DISTILLED OR DEIONIZED 
WATER.

• Inspect the electrode for scratches or

cracks. If any present, replace the elec-
trode.

• Cleaning procedure: clean frequently

the probe by soaking it for 1 minute in

HI70670

 or 

HI70671

 cleaning solu-

tions. After performing this procedure,
soak the electrode in 

HI70300

 storage

solution before taking measurements.

NOTE

For additional information about 
calibration, connections and 
readings, please refer to the instru-
ment (

HI9828

 multiparameter) 

instruction manual.