Dear Customer, 

Thank you for choosing a Hanna Instruments product. Please read this 
instruction manual carefully before using the Checker

®

HC handheld 

colorimeter. For more information about Hanna Instruments and our 
products, visit www.hannainst.com or e-mail us at [email protected].
For technical support, contact your local Hanna Instruments Office or e-mail 
us at [email protected].

Preliminary Examination

Remove the Checker

®

HC handheld colorimeter and accessories from the packing 

material and examine it carefully. If you require any further information, please 
contact Hanna Instruments technical support team.
Each 

HI764

 is delivered in a case with custom insert and is supplied with:

• Sample cuvette and cap (2 pcs.)
• Marine Nitrite Ultra Low Range reagent starter kit (reagents for 6 tests)
• 1.5V AAA Alkaline battery (1 pc.)
• Instruction manual

Note: 

Save all packing material until you are sure that the Checker

®

HC 

handheld colorimeter works correctly. Any damaged or defective item must 

be returned in its original packing material with the supplied accessories.

General Description & Intended Use

HI764

 Marine Nitrite Ultra Low Range handheld colorimeter is designed to 

determine nitrite levels in applications such as analysis of aquarium waters or 
analysis of wastewater and water quality.

HI764

 features a single-button operation system and is easy to use.

The large LCD is easy to read and the auto shut-off feature assures the battery 
will not be drained.

Specifications

Range

0 to 200 ppb (as NO

2

-N)

Resolution

1 ppb

Accuracy

±10 ppb ±4% of reading @ 25 °C (77 °F)

Light source

Light Emitting Diode @ 525 nm

Light detector Silicon photocell

Method

Adaptation of the EPA Diazotization Method 354.1. The 
reaction between nitrite and the reagent causes a pink tint 
in the sample.

Environment

0 to 50 °C (32 to 122 °F); max. 95% RH non-condensing

Battery type

1.5V AAA Alkaline

Auto shut-off After 10 minutes of non-use and 2 minutes after reading

Dimensions

86.0 x 61.0 x 37.5 mm (3.4 x 2.4 x 1.5”)

Weight

64 g (2.3 oz)

Functional Description & LCD Display

Cap
Cuvette with cap

Cuvette holder

LCD (Liquid Crystal Display)

ON/OFF button

Measurement Procedure

• Press the ON/OFF button to turn the checker 

on. All segments will be displayed for a few 
seconds, followed by “Add”

“C.1” with “Press” 

blinking.

• Fill the cuvette with 10 mL of unreacted sample 

and replace the cap. Insert the cuvette into the 
checker and close the cap.

• Press the ON/OFF button. When the display 

shows “Add”

“C.2” with “Press” blinking, the 

checker is zeroed.

• Remove the cuvette, unscrew the cap and add 

the content of one packet of 

HI764-0

 Nitrite 

Ultra Low Range reagent. Replace the cap and 
shake gently for about 15 seconds.

• Insert the cuvette into the checker and close the 

cap. Press and hold the ON/OFF button. The 
display will show the countdown prior to the 
measurement. 

• The instrument displays the nitrogen-nitrite 

(NO

2

-N) concentration in ppb. To convert 

the NO

2

-N concentration to the nitrite ion 

concentration (NO

2

), multiply the reading by 

factor of 3.29. The checker automatically turns 
off 2 minutes after reading.

Errors & Warnings

The checker shows clear warning messages when erroneous conditions appear and 
when measured values are outside the expected range. The information below 
provides an explanation of the errors and warnings, and the recommended action 
to be taken.

Light High:

 There is an excess amount of ambient light 

reaching the detector. Please check the preparation of the zero 
cuvette.

Light Low:

 There is not enough light to perform a measurement. 

Please check the preparation of the zero cuvette.

Inverted Cuvettes:

 The sample and the zero cuvettes are 

inverted. Swap the cuvettes and repeat the measurement.

Under Range:

 A blinking “0”

 

indicates that the sample absorbs 

less light than the zero reference. Check the measurement 
procedure and make sure to use the same cuvette for reference 
(zero) and measurement.

Over Range:

 Maximum concentration value displayed blinking 

indicates the measured value is outside the limits of the method.
Verify that the sample does not contain any debris. Dilute the 
sample and repeat the measurement.

Battery Low:

 Battery level is too low for the checker to function 

properly. Replace the battery with a new one.

Drained Battery:

 The battery is drained and must be replaced. 

Replace the battery with a new one and restart the checker.

Содержание HI764

Страница 1: ...HI764 Marine Nitrite Ultra Low Range INSTRUCTION MANUAL ...

Страница 2: ...is delivered in a case with custom insert and is supplied with Sample cuvette and cap 2 pcs Marine Nitrite Ultra Low Range reagent starter kit reagents for 6 tests 1 5V AAA Alkaline battery 1 pc Instruction manual Note Save all packing material until you are sure that the Checker HC handheld colorimeter works correctly Any damaged or defective item must be returned in its original packing material...

Страница 3: ... 354 1 The reaction between nitrite and the reagent causes a pink tint in the sample Environment 0 to 50 C 32 to 122 F max 95 RH non condensing Battery type 1 5V AAA Alkaline Auto shut off After 10 minutes of non use and 2 minutes after reading Dimensions 86 0 x 61 0 x 37 5 mm 3 4 x 2 4 x 1 5 Weight 64 g 2 3 oz Functional Description LCD Display Cap Cuvette with cap Cuvette holder LCD Liquid Cryst...

Страница 4: ...is zeroed Remove the cuvette unscrew the cap and add the content of one packet of HI764 0 Nitrite Ultra Low Range reagent Replace the cap and shake gently for about 15 seconds Insert the cuvette into the checker and close the cap Press and hold the ON OFF button The display will show the countdown prior to the measurement The instrument displays the nitrogen nitrite NO 2 N concentration in ppb To ...

Страница 5: ...d the zero cuvettes are inverted Swap the cuvettes and repeat the measurement UnderRange Ablinking 0 indicatesthatthesampleabsorbs less light than the zero reference Check the measurement procedure and make sure to use the same cuvette for reference zero and measurement Over Range Maximum concentration value displayed blinking indicates the measured value is outside the limits of the method Verify...

Страница 6: ...will be affected Discard the sample immediately after the reading has been taken or the glass might become permanently stained Battery Replacement To save the battery the checker shuts down after 10 minutes of non use and 2 minutes after reading A fresh battery lasts for a minimum of 5000 measurements When the battery is drained the instrument displays bAd then bAt and turns off To replace the bat...

Страница 7: ...onic Equipment The product should not be treated as household waste Instead hand it over to the appropriate collection point for the recycling of electrical and electronic equipment which will conserve natural resources Disposal of waste batteries This product contains batteries do not dispose of them with other household waste Hand them over to the appropriate collection point for recycling Ensur...

Страница 8: ...ce is not covered If service is required contact your local Hanna Instruments Office If under warranty report the model number date of purchase serial number and the nature of the problem If the repair is not covered by the warranty you will be notified of the charges incurred If the checker is to be returned to Hanna Instruments first obtain a Returned Goods Authorization RGA number from the Tech...

Отзывы: