•  Turn the meter on by pressing the button. All 

segments will be displayed. When the display 
shows “

Add

”, “

C.1

” with “

Press

” blinking, the 

meter is ready.

•  Rinse a graduated beaker several times with 

unreacted sample, before filling it to the  
50 mL mark with sample. 

•  Add 0.5 mL of 

HI719A-0

 reagent, swirl to mix.

•  Add 0.5 mL of 

HI719B-0

 reagent, swirl to mix.

•  Use this solution to rinse 2 cuvettes. Fill both 

cuvettes up to the 10 mL mark.

•  Add 1 drop of 

HI719C-0

 reagent to one cuvette 

(#1), replace the cap and invert the cuvette 
several times to mix. This is the blank.

•  Add 1 drop of 

HI719D-0

 reagent to the second 

cuvette (#2), replace the cap and invert the 
cuvette several times to mix. This is the sample.

•  Place the blank (#1) into the meter and close 

the meter’s cap.

•  Press the button. When the display shows 

Add

”, “

C.2

” with “

Press

” blinking the meter 

is zeroed.

•  Remove the cuvette.

•  Insert the second cuvette (#2) into the meter 

and press the button.

•  The meter directly displays the concentration 

of magnesium hardness (CaCO

3

) in ppm. The 

meter automatically turns off after 10 minutes.

#1

#2

#1

#2

#1

#2

Light High:

 There is too much light to perform a measurement. Please 

check the preparation of the zero cuvette.

Light Low:

 There is not enough light to perform a measurement. Please 

check the preparation of the zero cuvette.

Inverted Cuvettes:

 The sample and the zero cuvette are inverted.

Under Range:

 A blinking 

“0.00”

 indicates that the sample absorbs less 

light than the zero reference. Check the procedure and make sure you use 
the same cuvette for reference (zero) and measurement.

Over Range:

 A flashing value of the maximum concentration indicates 

the reading is over range. Dilute the sample and re-run the test.

Battery Low:

 The battery must be replaced soon.

Dead Battery:

 This indicates that the battery is dead and must be 

replaced. Once this indication is displayed, normal operation of the 
instrument will be interrupted. Change the battery and restart the meter.

Measurement Procedure

Hanna Instruments reserves the right to modify the design, construction or appearance of 

its products without advance notice.

Cover
Cuvette with cap

Cuvette holder

LCD Liquid Crystal Display

Operation Button

Functional Description

Errors and Warnings

Thank You

Thank you for choosing a Hanna Instruments product. Please read this instruction manual 
carefully before using the instrument.  
For more information about Hanna Instruments and our products, visit 
www.hannainst.com. 
For technical support, contact your local Hanna Instruments Office or e-mail us at tech@
hannainst.com. 
Find your local Hanna Instruments Office on www.hannainst.com.

Please examine this product carefully. Make sure that the instrument is not damaged. If 
any damage occurred during shipment, please contact your local Hanna Instruments Office.
Each 

HI719

 meter is supplied complete with:

• Sample Cuvettes and Caps (2 pcs.)
• Reagents set for 50 tests
• 100 mL Plastic Beaker (1 pc.)
• 1 mL Syringe with Tips (2 pcs.)
• 1.5V AAA Battery (1 pc.)
• Instruction Manual and Quick Reference Guide

Preliminary Examination

Range

0.00 to 2.00 ppm

Resolution

0.01 ppm

Accuracy

±0.20 ppm ±5% of reading @25 °C/77 °F

Light Source

Light Emitting Diode @525 nm

Light Detector

Silicon Photocell

Method

Adaptation of the Standard Methods for the Examination of Water 

and Wastewater, 18

th

 edition, EDTA colorimetric method. The reaction 

between magnesium and reagents causes a reddish-violet tint in the 

sample.

Environment

0 to 50 °C (32 to 122 °F); max 95% RH non-condesing

Battery Type

1.5V AAA (1 pc.)

Auto-Shut off

After 10 minutes of non-use

Dimensions

86.0 x 61.0 x 37.5 mm (3.4 x 2.4 x 1.5”)

Weight

52 g (1.84 oz.)

Specifications