14
Setup mode allows the selection of the Temperature unit,
Auto‑off, Beep, the type of pH buffer set, the Resolution and
Information. To enter Setup mode press and hold button
until “
STD
“ and
tag is replaced by “
SETUP
“ and
tag. Release the button.
• “
TEMP
“ is displayed on the
secondary LCD with the current
temperature unit (E.g. “
TEMP °C
“),
for °C/°F selection, use the
SET
button.
After the temperature unit has been
selected, press to confirm and to enter the “
A-OFF
“
selection.
• Use the
SET
button, to cycle
through the auto‑off choices:
8 minutes (
“8”
, default value), 60
minutes (
“60”
) or disabled (
“---”
).
Press to confirm and to enter the
“
BEEP
“ selection.
• To switch ON or OFF the beep
tone, press the
SET
button;
press to confirm and to enter
the calibration buffer selection
“
pH 7.01 BUFF
“.
• To change the type of calibration
buffer set, the meter will show
the current buffer set: “
pH 7.01
BUFF
” (for standard buffer set:
3.00/7.01/10.01) or “
pH 6.86
BUFF
” (for NIST buffer set: 3.00/6.86/9.18). Change
the set with the
SET
button. Press
to confirm and to
enter pH resolution selection “
RESOL
“.
• To change the pH measurement
resolution between “
0.1
“ and
“
0.01
“ use the
SET
button; then
press
to confirm and to enter
electrode calibration information
“
INFO
“ selection.
METER SETUP