Hanna Instruments HI 993310 Скачать руководство пользователя страница 10

10

B) Soil solution preparation and measurement:
1) Strain the soil with a 2-mm sieve;

2) Use 1 part of soil and 2 parts of distilled water;

3) Mix for 30 seconds;
4) Wait for 1 hour;
5) Filter the solution and measure the conductivity.

1) Liquid Crystal Display (LCD)

2) ON/OFF key

3) Water Conductivity selection key

4) Calibration Dial

5) Alarm LED

6) LED indicators

7) Soil Salinity selection key

ON

OFF

ALARM

S

C

-

+

HI 993310

mS

1

2

3

4

5

6

7

g/L

mS/cm

C WATER

Conductivity

S SOIL

Salinity

CALIBRATION

FUNCTIONAL DIAGRAM

FUNCTIONAL DIAGRAM

FUNCTIONAL DIAGRAM

FUNCTIONAL DIAGRAM

FUNCTIONAL DIAGRAM

11

SPECIFICATIONS

SPECIFICATIONS

SPECIFICATIONS

SPECIFICATIONS

SPECIFICATIONS

Range

g/L

0.00 to 1.00

mS/cm

0.00 to 19.99

Resolution

g/L

0.01

mS/cm

0.01

Accuracy

2% Full Scale

(@ 20°C / 68°F)

 from 0 to 15.00 mS/cm excluding probe error

Typical EMC

±2 % Full Scale

Deviation

Calibration

Water conductivity

Manual single setpoint through trimmer

Soil Salinity: Factory calibrated

Temperature

Automatic from 0 °C to 50 °C

Compensation

with a ß of 2% per degree °C

Probe

 (

included

)

   

HI 76305,amperometric, steel, ATC, 1m cable

HI 76304 amperometric, ATC, 1m cable

Both probes are included

Alarm

active when salinity value is lower than 0.20

and when salinity values is higher than

1.0

Environment

0 to 50°C (32 to 122°F);

max 95% RH non-condensing

Battery Type

 9 Volt (alkaline)

Life

100 hours of continuous use

Dimensions

185 x 82 x 45 mm (7.3 x 3.2 x1.8")

Weight

355 g (13 oz.)

Содержание HI 993310

Страница 1: ...NA LITERATURE HANNA LITERATURE Hanna publishes a wide range of catalogs and hand books for an equally wide range of applications The reference literature currently covers areas such as Water Treatment Process Swimming Pools Agriculture Food Laboratory Thermometry and many others New reference material is constantly being added to the library For these and other catalogs handbooks and leaflets cont...

Страница 2: ...chserv hannacan com These instruments are in compliance with the directives EN 50081 1 and EN 50082 1 19 CALIBRATION AND MAINTENANCE SOLUTIONS CHEMICAL TEST KITS CHLORINE METERS DISSOLVED OXYGEN METERS HYGROMETERS ION SPECIFIC METERS Colorimeters MAGNETIC STIRRERS Na NaCl METERS pH ORP Na ELECTRODES pH ORP METERS PROBES DO µS cm RH T TDS PUMPS REAGENTS SOFTWARE THERMOMETERS TITRATORS TRANSMITTERS ...

Страница 3: ...the packing material and examine it carefully to make sure that no damage has occurred during shipping If there is any noticeable damage notify your Dealer or the nearest Hanna office immediately Each meter is supplied with HI 76304 conductivity plastic probe HI 76305 soil stainless steel probe 9V battery Screwdriver Note Save all packing materials until you are sure that the instrument functions ...

Страница 4: ...il in a variety of applications among them FERTILAZATION By measuring the salt contents in soil and substrates before and after the fertilization make it possible to know the soil fertility and to devise an appropriate fertilization plan Consequently the plants will receive the right quantity of nutrients resulting in a regular and rapid growth greater resistance to disease and a more appealing st...

Страница 5: ...th deep roots e g trees maize sunflower etc it is advised to measure at different depths e g 30 50 and 80 cm that is 12 20 and 32 Take the measure in more spots to have a representative average The soil has to be wet If the soil is dry add demineralized water The tip of the probe must be in complete contact with the soil If the substrate is too soft press it with your fingers to achieve a proper c...

Страница 6: ...turist 15 BATTERY REPLACEMENT BATTERY REPLACEMENT BATTERY REPLACEMENT BATTERY REPLACEMENT BATTERY REPLACEMENT When the battery voltage is low the meter will display V When the low battery indicator appears the battery has only a few hours left HI 993310 HI 993310 HI 993310 HI 993310 HI 993310 will then automatically turn off itself to prevent erroneous measurements due to low battery It is recomme...

Страница 7: ...ration knob until the display shows the conductivity value of the calibration solutions at 25 C 12 88 mS cm All subsequent measurements will be compensated to 25 C 77 F The calibration is now complete and the instrument is ready for use The instrument should be recalibrated at least once a month or when the probe or battery is changed Note For more accurate results it is advisable to use a calibra...

Страница 8: ...y requiring lower concentrations in the summer Asparagus Watermelon Carrot Cabbage Cucumber Crysanthemum 1 50 2 00 1 50 2 50 1 50 2 00 2 00 3 00 2 00 3 00 1 50 2 50 PLANT PREFERED EC VALUES LIQUID PROBE LIQUID PROBE LIQUID PROBE LIQUID PROBE LIQUID PROBE 13 than 30 C allow more time for the thermal equilibrium of the system to be achieved The meter provides for an LED alarm when measuring soil sal...

Страница 9: ...ATION OF SAMPLE PREPARATION OF SAMPLE PREPARATION OF SAMPLE 12 OPERATIONAL GUIDE OPERATIONAL GUIDE OPERATIONAL GUIDE OPERATIONAL GUIDE OPERATIONAL GUIDE Each meter is supplied complete with a 9V battery and two probes Slide off the battery compartment cover on the back of the meter see page 15 Install the battery while paying attention to its polarity Connect the right probe for your measurement C...

Страница 10: ...NS SPECIFICATIONS SPECIFICATIONS SPECIFICATIONS Range g L 0 00 to 1 00 mS cm 0 00 to 19 99 Resolution g L 0 01 mS cm 0 01 Accuracy 2 Full Scale 20 C 68 F from 0 to 15 00 mS cm excluding probe error Typical EMC 2 Full Scale Deviation Calibration Water conductivity Manual single setpoint through trimmer Soil Salinity Factory calibrated Temperature Automatic from 0 C to 50 C Compensation with a ß of ...

Отзывы: