Make sure that the pH electrode and any ground probe have been
properly connected and wired to the controller (see preceding pages)
and that the meter is plugged to the mains.
Calibration should be performed at a tem-
perature similar to that of the liquid to be
monitored.
Use a Checktemp (or an accurate thermom-
eter) as reference. Remove the electrode cap if
it is still on the electrode.

During calibration, move the electrode and the
separate ground probe (if in use) together
from one buffer to the next.

If no separate ground probe is being used, make sure that
the Reference and the Matching Pin terminals are shorted
(see 14 - Functional Diagram).
If the electrode (such as HI 1003/3) incorporates a ground probe
(matching pin) then remove the jumper.

OFFSET ADJUSTMENT:

• Rinse the tip of the electrode with pH 7.01 solution (

HI 7007

),

then dip the bottom 4 cm (1.5”) of the
electrode (and ground probe) in the pH
7.01 buffer.

• Place the Checktemp thermometer in the

buffer solution.

• Wait for the measurement to sta-

bilize and then adjust the
"OFFSET" trimmer (see 6 - Func-
tional Diagram) to display
pH 7.01 on the LCD if the tem-
perature of the buffer solution is
at 25°C (77°F).

• If the temperature of the buffer solution is not 25°C (77°F), refer

to the chart at the end of the manual for the appropriate buffer
value at a given temperature.

pH 

pH 

pH 

pH 

pH  CALIBRATION

CALIBRATION

CALIBRATION

CALIBRATION

CALIBRATION

Содержание HI 9912

Страница 1: ...ves ALARM LINE SET ORP SET pH 0 SEC OND S mV 200 0 ORP TIMER MINUTE S 90 PRO POR TIONAL SET TINGS 1 10 2 5 5 7 5 0 SECO NDS pH 2 0 0 pH TIME R MINU TES 90 PROPORTION AL SETT INGS 1 10 2 5 5 7 5 100 50...

Страница 2: ...ation do not hesitate to e mail us at tech hannainst com These instruments are in compliance with the directives PRELIMINARY EXAMINATION 3 GENERAL DESCRIPTION 3 MECHANICAL LAYOUTS 4 FUNCTIONAL DIAGRAM...

Страница 3: ...strument s main power line as well as the dosage and alarm terminals are protected by separate 2A fuses Use only 2A fuses for replacement Hanna s wall mounted pH and ORP controller with proportional c...

Страница 4: ...IN Fig 2 Figure 2 illustrates the controls and terminals MECHANICAL LAYOUTS MECHANICAL LAYOUTS MECHANICAL LAYOUTS MECHANICAL LAYOUTS MECHANICAL LAYOUTS ELECTRODE WITH BNC CONNECTION WIRING ACCESS PORT...

Страница 5: ...then close the compartment Adjustments can then be made only in the control area without having to open the contacts compartment Figure 4 is a dimensioned front view of the Wall Mounted Controller Th...

Страница 6: ...for pH 2 Liquid Crystal Displays for pH and ORP 3 Acid feed LED and dial to adjust the pH setpoint 4 Alarm disable switches for pH and ORP 5 Slope calibration trimmer for pH 6 Offset calibration trim...

Страница 7: ...ed dosage terminals for pH correction 19 Powered dosage terminals for ORP correction 20 110 115V or 220 240V power configuration 21 Incoming power terminals 22 Fuses BOTTOM VIEW 23 Female BNC connecto...

Страница 8: ...to 200 mV with two separate time cycle from 0 to 90 seconds ALARM Relay can be configured as normally open or normally CONTACT closed isolated output Max 2A Max 240V resistive load 1 000 000 strokes T...

Страница 9: ...r replace the cover replace the cover replace the cover Only then connect the controller to the power supply ELECTRODE GROUND PROBE CONNECTIONS Simply attach any combination pH and ORP electrode with...

Страница 10: ...as normally closed NC by connecting the external device to the C and NC terminals or normally open NO by connecting the external device to the C and NO terminals RELAY CONNECTIONS Wire the external d...

Страница 11: ...minal is activated triggering the mechanism of your choice The alarm LED lights also come on Due to the Consent Consent Consent Consent Consent feature feature feature feature feature if the pH is in...

Страница 12: ...nd another for ORP The LED s light up when the controller is in the pH or ORP dosage mode and the terminals are activated 500 900 SET FEED mV ALARM OXID 700 pH ALARM 8 SET FEED 6 7 ACID For 110 115V s...

Страница 13: ...e that the Reference and the Matching Pin terminals are shorted see 14 Functional Diagram If the electrode such as HI 1003 3 incorporates a ground probe matching pin then remove the jumper OFFSET ADJU...

Страница 14: ...e and any ground probe have been properly connected and wired to the controller see preceding pages and that the meter is plugged to the mains Remove the electrode cap if it is still on the electrode...

Страница 15: ...ONAL SETTINGS 0 5 1 5 0 SECONDS mV 200 0 90 PROPORTIONAL SETTINGS 50 150 PROPORTIONAL CONTROL PROPORTIONAL CONTROL PROPORTIONAL CONTROL PROPORTIONAL CONTROL PROPORTIONAL CONTROL In order to optimize t...

Страница 16: ...responding samples or highly concentrated chemicals or under dosage with slow reacting or weak chemicals the controller provides even a more accurate control As the graphs show it does that by stoppin...

Страница 17: ...ce acti vated for 15 seconds and off for 45 seconds until the next cycle 0 SECONDS mV 200 0 90 PROPORTIONAL SETTINGS 50 150 OVERDOSAGE TIMERS OVERDOSAGE TIMERS OVERDOSAGE TIMERS OVERDOSAGE TIMERS OVER...

Страница 18: ...IOUS VALUES AT VARIOUS TEMPERATURES TEMPERATURES TEMPERATURES TEMPERATURES TEMPERATURES TEMP pHVALUES C F 4 01 6 86 7 01 9 18 10 01 0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 32 41 50 59 68 77 86 95 1...

Страница 19: ...g mediums This phenomenon is rough on the electrode To make accurate redox measurements the surface of the electrode must be clean and smooth At certain mV and pH values especially those found in pool...

Страница 20: ...these bubbles If the bulb and or junction are dry soak the electrode in a HI 70300 Storage Solution overnight STORAGE To minimize clogging and assure a quick response time the glass bulb and the junc...

Страница 21: ...e tip in Hanna Solu tion HI 7061 for 30 minutes rinse thoroughly with distilled water and then recalibrate the meter For ORP Electrodes Polish the metal tip with a lightly abrasive paper paying attent...

Страница 22: ...ellent insulation even though Amphel elec trodes can operate with both AmpHel electrodes have a micro amplifier in the electrode cap to boost the signal drastically reducing susceptibility to noise an...

Страница 23: ...ow rates from 1 5 to 18 3 lph 0 4 to 4 8 gph ChecktempC Pocket size thermometer range 50 0 to 150 0 C ChecktempF Pocket size thermometer range 58 0 to 302 0 F HI 6050 Submersible electrode holder 605...

Страница 24: ...24...

Страница 25: ...y the warranty you will be notified of the charges incurred If the instrument is to be returned to Hanna Instruments first obtain a Returned Goods Authorization Number from the Customer Service depart...

Страница 26: ...ITTERS TURBIDITY METERS Wide Range of ACCESSORIES Most Hanna meters are available in the following formats BENCH TOP METERS POCKET SIZED METERS PORTABLE METERS PRINTING LOGGING METERS PROCESS METERS P...

Страница 27: ...ent Radiated Class B User Safety Requirement Date of Issue 07 06 1999 D Volpato Engineering Manager On behalf of Hanna Instruments S r l IEC 801 2 IEC 801 3 IEC 801 4 EN 55022 EN 61010 1 Recommendatio...

Страница 28: ...cations The reference literature currently covers areas such as Water Treatment Process Swimming Pools Agriculture Food Laboratory Thermometry and many others New reference material is constantly bein...

Отзывы: