INITIAL PREPARATION

Each meter is supplied complete with batteries. Remove the back
cover, unwrap the batteries and install them while paying attention to
the polarity. Alternatively, connect the HI 710005 or HI 710006
voltage adapter to the power adapter plug.
To prepare the instrument for use, choose the most appropriate tem-
perature probe for your application (see accessories) and connect it to
the connector located on the top of the instrument.
All the probes have been factory precalibrated and no calibration is
needed.
To turn the meter on, press the ON/OFF key.
The batteries charge status or "LINE" message
(if external power adapter is connected) will be
displayed on the LCD for a few seconds.
The meter is now ready to operate.

To maximize battery life, the meter is automatically turned off after a
user selectable  period of non-use (this feature is enabled and set to
5 minutes by default; it can be disabled or changed through setup
code 40). If in logging mode, after the period of non-use, the meter
will continue to monitor the temperature periodically at the end of
every logging interval. Only the "LOG" indication will be visible on
LCD. While storing data in memory, during the sleep mode, the
reading will appear briefly on the LCD.
To reactivate the display, press the 

ON/OFF

 key.

Note:

When the use of an alternate function (FNC, CFM and LOT
INFO) is requested, press and hold the 

ALT

 key first and then

the second listed key.

FNC

Содержание HI 98811

Страница 1: ...Instruction Manual HI 98811 Fast Tracker Thermometer www hannainst com...

Страница 2: ...ROBES 24 UN HOUSED NTC SENSORS 25 ACCESSORIES 26 TABLE OF CONTENTS HI 98811 is guaranteed for two years against defects in workmanship and materials when used for its intended purpose and maintained a...

Страница 3: ...made of rugged and lightweight material making it truly portable The meter comes equipped with an easy to read LCD with backlight feature for comfortable reading even in dark environments A user frien...

Страница 4: ...m field and lab applications this meter can be connected to a 12VDC adapter HI 98811 has the capability to store the measurements in memory at a user selectable interval from 1 to 180 minutes This inf...

Страница 5: ...key to select temperature reading to view date and time and to enable backlight with ALT 8 LOG key to store and or print measurements 9 LOT INFO key to move down or view logging information with ALT 1...

Страница 6: ...inter Low power impact type belt 14 characters per line 38 mm plain paper HI 710034 Printing Logging 1 2 5 10 15 30 60 120 and 180 minutes Interval Serial interface Infrared RS232 selectable at 1200 2...

Страница 7: ...external power adapter is connected will be displayed on the LCD for a few seconds The meter is now ready to operate To maximize battery life the meter is automatically turned off after a user selecta...

Страница 8: ...in access to stored data To enter setup mode ensure the meter is not logging and then press the ALT and FNC keys The scrolling message Insert the function code or press ALT FNC to escape on the upper...

Страница 9: ...message will scroll across the LCD for a few seconds The current value of the selected parameter on the upper LCD and the parameter code on the lower LCD will be displayed The first digit will blink i...

Страница 10: ...mm 59 01 dd 28 29 30 31 1 MM 12 1998 YYYY 2097 2 To change the password the correct code must be entered first If the password has been forgotten the password protected features are no longer accessi...

Страница 11: ...ice across the LCD The upper LCD will then display L 00 with the 00 blinking Set the desired lot using the Arrow keys L 00 is the lot of data of the log on demand and L 01 to L 16 are the lots of the...

Страница 12: ...le Data will be displayed in the following order Sample ID for Lot 00 only temperature value date time If TEMP TIME is pressed when the time is displayed the LCD will pass to the next sample number It...

Страница 13: ...an be viewed in reverse order pressing the key Press ALT and FNC to return to function selection mode Code 20 setting the instrument identification ID code When using several identical meters it may b...

Страница 14: ...view the date and time in the following order date time Pressing TEMP TIME again the meter re turns to temperature reading If the reading is out of range or the probe is not connected the LCD will dis...

Страница 15: ...ll be displayed on the upper part of the LCD and Id on the lower part For Sample ID code reading an i Button tag must be touched by the touch probe on the back of the meter within a 20 second inter va...

Страница 16: ...g and printing make sure there is enough paper for your measurements When the paper is finished the meter will not advise the operator and the printouts could be lost If this happens data will continu...

Страница 17: ...ed Note Tag reading timeout after 20 seconds left without reading any tag the logging operation will be aborted The LCD will display Stor and the value will be stored in the lot 00 log on demand data...

Страница 18: ...lot with the and keys Press to view older lots or to view more recent ones If is pressed when the last lot is displayed the meter displays the lot 00 log on demand By pressing the key again the meter...

Страница 19: ...p lithium battery This allows the meter to update time and date even when both batteries and external power adapter are disconnected GOOD LABORATORY PRACTICE GLP GLP is a set of functions that allows...

Страница 20: ...to your PC for easy reproduction storage or elaboration without the need of cables between the meter and the transmitter Data transmission from the instrument to the PC is supported by the HI 92000 W...

Страница 21: ...ension of the old lot are updated Note The oldest lot data are erased without any warning message Note Timed logging memory can be entirely erased through the setup code 06 If the meter is powered onl...

Страница 22: ...ace the paper cover The printing logging thermometer is factory programmed to auto matically diagnose a fault and to display error codes on the LCD PRINTER ERROR Whenever a printer fault condition is...

Страница 23: ...light and printer when the instrument is in a low battery condition If the user attempts to enable these functions without replacing the batteries or connecting the external power adapter the meter wi...

Страница 24: ...able blue handle HI762PBL 10 Penetration probe 10 m 33 cable blue handle HI762PG Penetration probe with 1 m 3 3 cable green handle HI762PG 10 Penetration probe 10 m 33 cable green handle HI762PR Penet...

Страница 25: ...2 1 NTC Thermistor sensor Teflon coated with hole for mounting HI 76S2 2 NTC Thermistor sensor Self adhesive foam disk for mounting HI 76S2 3 NTC Thermistor sensor Eye connector for mounting HI 76S2 4...

Страница 26: ...6 4mm 0 25 25 4mm 1 0 300mm 12 15 8mm 0 625 7 1mm 0 28 MOLDED PVC TIP 2 4mm 0 095 LEAD DIA 0 2mm 0 008 38mm 1 50 Windows is registered Trademark of Microsoft Co HI710005 Voltage adapter from 115 VAC t...

Страница 27: ...ow all necessary steps to correct interferences Any variation introduced by the user to the supplied equipment may degrade the instrument s EMC performance To avoid electrical shock do not use this in...

Страница 28: ...9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 011 615 8582 Taiwan Tel 886 2 2739 3014...

Отзывы: