OPERATION

• Do not be alarmed if white crystals appear around

the cap. This is normal with pH electrodes and
they dissolve when rinsed with water.

• Take a sample of the water to be tested and pour

into the black compartment of the Water Test.
If you are scooping up the water with the black
compartment, do not immerse above the indicated
maximum immersion level.

• Turn the tester on by pressing the ON/OFF

key on the front panel.

• The tester will enter the temperature mea-

surement mode, indicated by the "°C" sym-
bol on the display.

• Press the RANGE key once to enter the

conductivity (µS/cm) mode.

• Press the RANGE key again to display pH.
• Press the RANGE key once again to dis-

play ORP (mV).

• If the tester compartment has been left dry

for more than 2 days, wait for about 5 min-
utes before taking pH measurements.

• For more accurate measurements, make sure that Water Test is

calibrated for pH and conductivity before use.

• After measurements, empty the compartment, rinse it with tap water

and switch the tester off by pressing the ON/OFF key.

• When the tester is not in use, fill the compartment

with tap water.
For long term storage, keep the tester in its origi-
nal box.
NEVER USE DISTILLED OR DEIONIZED WATER
FOR STORAGE PURPOSES.

pH CALIBRATION

• Pour a small quantity of distilled water into the

black compartment, rinse by shaking the compart-
ment and then discard the content.

• Pour a small quantity of pH 7 (HI 7007) buffer solu-

tion into the compartment and switch the tester on
by pressing the ON/OFF key.

• Press the RANGE key until the display enters the

pH measurement mode.

• Adjust the right trimmer located behind

the front cover until the display shows "7.0
pH".

• Discard the pH 7 solution and rinse the

compartment with distilled water.

• Pour a small quantity of pH 4 (HI 7004) or

pH 10 (HI 7010) buffer solution into the com-
partment.

• Adjust the left  trimmer located behind the

front cover until the display shows "4.0 pH"
or "10.0 pH".

• Discard the buffer solution and rinse the

compartment with distilled water.

• The unit is now calibrated for pH and ready

for use.

CONDUCTIVITY CALIBRATION

• Rinse the black compartment with distilled

water, then fill it with HI 7031 conductivity so-
lution to overflow and switch the tester on by
pressing the ON/OFF key. The temperature
of the solution will be displayed.

• Note the temperature of the solution and read

the corresponding conductivity from the table
on the solution bottle label (e.g. 1413  µS/cm
@25°C).

• Remove the screw on the right side of the casing to access the

conductivity calibration trimmer. Press the RANGE key to display
the conductivity measurement mode and adjust the trimmer
until the display shows the conductivity reading at the noted tem-
perature (e.g. 1278 µS/cm @20°C). Replace the screw after calibra-
tion.

°C

°F

HI7031

S/cm)

0

32

776

5

41

896

10

50

1020

15

59

1147

16

60.8

1173

17

62.6

1199

18

64.4

1225

19

66.2

1251

20

68

1278

21

69.8

1305

°C

°F

HI7031

S/cm)

22

71.6

1332

23

73.4

1359

24

75.2

1386

25

77

1413

26

78.8

1440

27

80.6

1467

28

82.4

1494

29

84.2

1521

30

86

1548

31

87.8

1575

ALWAYS USE FRESH BUFFERS

FOR CALIBRATION