Working with custom buffers

• To select a custom buffer (see Note): press the 

ARROW

 keys while in pH calibration

mode until the 

Custom

 key will be displayed on the LCD.

• To change the custom buffer value in accordance with current temperature:

press 

Custom

 and then the 

ARROW

 keys. Press 

Accept

 to confirm the value.

Note:

The custom buffer is supposed to be set in SETUP mode.

Replacing an old calibration point

• To select the old calibration point to be replaced: press the 

ARROW

 keys.

• To confirm the old calibration point to be replaced: press 

CFM

.

• To leave replace mode: press 

CAL

.

pH Calibration messages:

Wrong buffer

 - the pH reading is far from the selected buffer.

Electrode dirty/broken

 alternatively with  “

Buffer Contaminated

” the electrode

offset is out of acceptable range.

Wrong Buffer

 

Temperature

: buffer temperature out of defined range.

Wrong Old Slope

: inconsistency between current and old calibration parameter.

Clean Electrode

: an abnormal difference between new and previous calibration

detected.

Outside Cal Range

: out of calibration range.

RELATIVE mV CALIBRATION PROCEDURE

• To enter/exit relative mV calibration mode: press 

CAL

 while in relative mV

measurement mode; the relative mV value is displayed on the main LCD line and
the absolute mV value on the secondary line.

• To change the displayed relative mV value: press the 

ARROW

 keys.

• To confirm relative mV calibration: press 

CFM

. The meter returns to relative mV

measurement mode.

ION CALIBRATION PROCEDURE

• Place the ISE probe in the less concentrated standard.
• Wait for “

CFM

” to blink. Press 

CFM

 to accept the calibration point (see Note).

• Follow the above procedure until the fifth calibration point is reached. The meter

returns to measurement mode.

Note:

For one, two, three or four-point calibration, press 

CAL

 after the appropriate

accepted solution. The meter returns to measurement mode.

SETUP MODE

• To enter/exit SETUP view mode: press 

SETUP

 while in measurement mode.

• To toggle between displayed parameters: press the 

ARROW

 keys.

• To enter/exit SETUP changing mode: press 

Modify

 or simply select option.

GLP MODE

• To enter/exit GLP mode: press 

GLP

 while in measurement mode.

LOG MODE

• To store the current reading in memory: press 

LOG

 while in measurement mode.

• To view logged data: press 

RCL

 while in measurement mode.

• To view more details: press 

More

.

• To delete records: press 

Delete

 key.

• To delete all records: press 

Delete All

.

AutoEnd MODE

• To freeze the first stable reading on the LCD: press 

AutoEnd

 while in measurement

mode; the “

WAIT

” tag will blink until the reading will stabilize, then “

Hold

will be displayed.

• To return to measurement mode: press 

Continue

.