OPERATIONAL GUIDE

TAKING MEASUREMENTS

• Turn the meter on.
• Immerse the tip (5 cm/2”) of

the probe in the sample.

Note:

 For better accuracy, allow

the reading to stabilize. Any
initial variation may be due
to temperature compensa-
tion and the fact that the
probe is adjusting itself to
the new sample.

For better accuracy, a frequent calibration is recommended.
Moreover, the instrument should be recalibrated at least once
a month, after cleaning the probe or replacing the batteries.
• Turn the meter on.
• Press and hold the ON/OFF

button for approx. 3 seconds.
The display will start blinking
“1500 ppm” for 

HI 981409

and “5.00 ms” for 

HI 981410

to indicate that the calibra-
tion mode has been entered.

• Immerse the probe in the proper cali-

bration solution: use 

HI 70442

 (1500

ppm) for 

HI 981409

, or 

HI 70039

(5.00 mS/cm) for 

HI 981410

.

• Stir gently and wait for approximately

20 seconds for the meter to perform
the automatic calibration.

Note:

 The probe should be submerged

approximately 5 cm (2”) in the
solution.

• Once the display stops blinking, the

instrument is calibrated and ready to
use.

CALIBRATION

ON/OFF

1500 ppm

BUFFER SOLUTION

20 mL sacket

HI 70442P

1500

TDS

ppm

ACCESSORIES

HI 1220

Spare probe for HI 981409 and HI 981410,
240 mm (9.4”) long, with DIN connector and
built-in temperature sensor

HI 70039P

5.00 mS/cm calibration solution, 20 mL sachet
(25 pcs)

HI 70442P

1500 ppm calibration solution, 20 mL sachet
(25 pcs)

HI 7039M

5.00 mS/cm calibration solution, 230 mL bottle

HI 70442M

1500 ppm calibration solution, 230 mL bottle

HI 7061M

Probe cleaning solution, 230 mL bottle

CE DECLARATION OF CONFORMITY

DECLARATION OF CONFORMITY

We

Hanna Instruments Srl
Via E.Fermi, 10
35030 Sarmeola di Rubano (PD)
ITALY

herewith certify that the EC / TDS meter

has been tested and found to be in compliance with the following regulations:

Electrostatic Discharge
RF Radiated
Fast Transient
Radiated, Class B

Date of Issue: 26-10-1999

D.Volpato - Engineering Manager

On behalf of

Hanna Instruments S.r.l.

HI 981409 HI 981410

IEC 801-2
IEC 801-3
IEC 801-4
EN 55022
EN 61010-1

Electrical Safety

Recommendations for Users 

Before using this product, make sure that it is entirely suitable for the environment in
which it is used. Operation of this instruments in residential areas could cause unacceptable
interference to radio and TV equipment.
The metal band at the end of the probe is sensitive to electrostatic discharges. Avoid touching
this metal band at all times. During operation, ESD wrist straps should be worn to avoid possible
damage to the probe by electrostatic discharges.
Any variation introduced by the user to the supplied equipment may degrade the instrument’s
EMC performance. To avoid electrical shock, do not use this instrument  when voltage at the
measurement surface exceed 24 Vac or 60 Vdc.
Use plastic beakers to minimize any EMC interferences.
To avoid damage or burn, do not perform any measurement in microwave ovens.

IST981409R1     07/05

ERROR CODES

BATTERY REPLACEMENT

When the batteries are weak, EC and TDS 

 

will

display the warning symbol “Eb”.
To replace the batteries, remove the rear panel and replace
both batteries, while paying attention to their polarity.
Batteries should only be replaced in a safe area and using
1.5V AA alkaline batteries.

Note

: Once the batteries have been replaced, turn the meter

OFF and then ON again. Wait for 30 seconds to allow
the meter to complete the selftest before performing a
new calibration.

Error in calibration, wrong buffer solution
or probe requiring cleaning or replacing.

Over range

Low battery indication

ON/OFF

5.00

EC

PROBE CONNECTION & MAINTENANCE

In order to protect the DIN connector against
vapors and humidity, the meter is provided
with a waterproof sheath.

• Slide the protective sheath down, connect the 

HI 1220

probe to the DIN connector and then slide the protective
sheath back up. To ensure maximum protection, make
sure the connector is completely covered.

• If the meter is not used for a long time, it is recommended

to clean the internal sensitive pins with alcohol. Immerse
the tip of the probe up to the air-vent hole in alcohol for
a few minutes.

• After use, clean the probe with tap water and dry it.
• After prolonged use, clean the probe by immersing it in

HI 7061

 cleaning solution for 30 minutes.