Accessories needed:

• Use 

HI 70031

 (1413 

μ

S/cm = 1.413 mS/cm) EC calibration

solution or 

HI 70442

 (1500 ppm) TDS calibration solution.

Note:

The conversion between EC and TDS is made by a built-in
circuit, hence it is requested to calibrate the meter only in EC
or TDS range. The other range is thus automatically calibrated.

PROCEDURE

• Pour approximately 4 cm (1½") of a conductivity calibration

solution (e.g. 

HI 7031

) into a beaker. If possible, use plastic

beaker to minimize any EMC interference.

• Immerse the probe in the solution.
• Wait for a couple of minutes for thermal equilibrium to be

reached.

• Tap the probe on the bottom, then shake it lightly while rotating

to make sure no air bubbles remain trapped inside the probe.

• Press the 

mS/cm

 (or 

ppm

) key.

• Turn the EC/TDS calibration knob until the

display shows the EC or TDS reading at
25°C.

EC/TDS

EC/TDS

EC/TDS

EC/TDS

EC/TDS CALIBRATION

 CALIBRATION

 CALIBRATION

 CALIBRATION

 CALIBRATION

Содержание HI 9813-6

Страница 1: ...1 Instruction Manual HI 9813 5 HI 9813 6 Portable pH EC TDS C Meters www hannainst com...

Страница 2: ...BLE OF CONTENTS TABLE OF CONTENTS HI 9813 5 and HI 9813 6 are guaranteed for two years against defects in workmanship and materials when used for their intended purpose and maintained according to ins...

Страница 3: ...ake sure that no damage has occurred during shipment If noticeable damage is found notify your dealer Each meter is supplied with Combination probe with 8 pin DIN connector 1 m 3 3 cable HI 1285 5 pH...

Страница 4: ...on key 5 ppm TDS selection key 6 pH offset calibration knob 7 EC TDS calibration knob 8 C Temperature selection key 9 On Off key FUNCTIONAL DESCRIPTION OF HI 9813 5 FUNCTIONAL DESCRIPTION OF HI 9813 5...

Страница 5: ...pm TDS range selection key 6 pH offset calibration knob 7 EC TDS calibration knob 8 C Temperature selection key 9 Check key 10 On Off key FUNCTIONAL DESCRIPTION OF HI 9813 6 FUNCTIONAL DESCRIPTION OF...

Страница 6: ...a r b i l a C S D T C E e p o l s h g u o r h t t n i o p 1 l a u n a M r e m m i r t e r u t a r e p m e T S D T C E n o i t a s n e p m o C F 2 2 1 o t 2 3 C 0 5 o t 0 m o r f c i t a m o t u A h t...

Страница 7: ...simply submerge the tip 4cm 1 of the probe into the sample to be tested Select the pH mode Stir briefly and wait a couple of minutes for the reading to adjust and stabilize The display shows the pH v...

Страница 8: ...S mode TAKING TEMPERATURE MEASUREMENTS Submerge the tip 4cm 1 of the probe into the sample to be tested Select the C mode Stir briefly and wait a couple of minutes for the reading to adjust and stabil...

Страница 9: ...HI 7009 instead of pH 10 01 PROCEDURE Connect the probe and switch the meter on then press the pH key to display pH measurement Remove the protective cap from the probe rinse and immerse it in the bu...

Страница 10: ...4 0 9 5 8 9 0 5 2 2 1 6 0 4 3 8 6 8 9 6 1 0 9 2 8 9 5 5 1 3 1 8 0 4 4 8 6 8 9 6 9 9 8 9 7 9 0 6 0 4 1 9 0 4 4 8 6 8 9 6 7 9 8 7 7 9 5 6 9 4 1 1 1 4 4 8 6 9 9 6 5 9 8 6 7 9 0 7 8 5 1 2 1 4 5 8 6 9 9 6...

Страница 11: ...nductivity calibration solution e g HI 7031 into a beaker If possible use plastic beaker to minimize any EMC interference Immerse the probe in the solution Wait for a couple of minutes for thermal equ...

Страница 12: ...dous area using a 9V alkaline battery Unscrew the three screws on the rear of the meter remove the battery compartment cover and replace the 9V battery with a new one while observing its polarity Make...

Страница 13: ...be is broken and has to be replaced Press the Check key If the meter is calibrated Probe is OK message is displayed If cleaning is needed Clean Probe and Calibrate message is displayed 1 immerse the p...

Страница 14: ...ion 500 mL bottle HI 7010L pH 10 01 buffer solution 500 mL bottle CONDUCTIVITY TDS CALIBRATION SOLUTIONS HI 7031L 1413 S cm solution 500 mL bottle HI 7442L 1500 ppm mg L solution 500 mL bottle OTHER S...

Страница 15: ...follow all necessary steps to correct interferences The glass bulb at the end of the electrode is sensitive to electrostatic discharges Avoid touching this glass bulb at all times During operation ES...

Страница 16: ...Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA Technical Support for Customers Tel 800 426 6287 Fax 401 765 7575 E mail tech hannainst com www hannainst com...

Отзывы: