PROBE MAINTENANCE

These meters are powered by a 9V battery that is located on the rear 
of the instrument.
When battery symbol is empty a low battery condition is indicated.
When the low battery indication appears, only a few hours of working 
time remains.

It is recommended to replace the battery immediately.
When the battery level is so low that it may cause unreliable measure-
ments, the meter turns off.
Battery replacement must only take place in a nonhazardous area using 
a 9V alkaline battery.
Unscrew the three screws on the rear of the meter, remove the battery 
compartment cover and replace the 9V battery with a new one while 
observing its polarity.
Make sure the battery contacts are tight and secure before replacing 
the cover.

BATTERY REPLACEMENT

PERIODIC MAINTENANCE

Inspect the probe and the cable. The cable used for the connection to the 
meter must be intact and there must be no points of broken insulation 
on the cable or cracks on the probe stem or bulb.
Connector must be perfectly clean and dry. If any scratches or cracks are 
present, replace the electrode. Rinse off any salt deposits with water.

Содержание HI 9813-5

Страница 1: ...1 Instruction Manual HI9813 5 HI9813 6 Portable pH EC TDS C Meters www hannainst com...

Страница 2: ...5 and HI9813 6 are guaranteed for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions Electrodes and probes are guarant...

Страница 3: ...emove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipment If noticeable damage is found please contact your local Hanna Instrumen...

Страница 4: ...Liquid Crystal Display 3 pH range selection key 4 mS cm EC range selection key 5 ppm TDS selection key 6 pH offset calibration knob 7 EC TDS calibration knob 8 C Temperature selection key 9 On Off key...

Страница 5: ...ystal Display 3 pH range selection key 4 mS cm EC range selection key 5 ppm TDS range selection key 6 pH offset calibration knob 7 EC TDS calibration knob 8 C Temperature selection key 9 Check key 10...

Страница 6: ...onversion Factor variable 0 56 to 0 78 1 S cm according with TDS442 curve pH Calibration Manual 1 point through offset trimmer EC TDS Calibration Manual 1 point through slope trimmer EC TDS Tempera tu...

Страница 7: ...0300 storage solution for 30 minutes to reactivate it To take a pH measurement simply submerge the tip 4 cm 1 of the probe into the sample to be tested Select the pH mode Stir briefly and wait a coupl...

Страница 8: ...S mode TAKING TEMPERATURE MEASUREMENTS Submerge the tip 4 cm 1 of the probe into the sample to be tested Select the C mode Stir briefly and wait a couple of minutes for the reading to adjust and stabi...

Страница 9: ...tead of pH7 01 and pH9 18 HI7009 instead of pH10 01 PROCEDURE Connect the probe and switch the meter on then press the pH key to display pH measurement Remove the protective cap from the probe rinse a...

Страница 10: ...C F 4 01 6 86 7 01 9 18 10 01 0 32 4 01 6 98 7 13 9 46 10 32 5 41 4 00 6 95 7 10 9 39 10 24 10 50 4 00 6 92 7 07 9 33 10 18 15 59 4 00 6 90 7 05 9 27 10 12 20 68 4 00 6 88 7 03 9 22 10 06 25 77 4 01...

Страница 11: ...DURE Pour approximately 4cm 1 of a conductivity calibration solution e g HI7031 into a beaker If possible use plastic beaker to minimize any EMC interference Immerse the probe in the solution Wait for...

Страница 12: ...azardous area using a 9V alkaline battery Unscrew the three screws on the rear of the meter remove the battery compartment cover and replace the 9V battery with a new one while observing its polarity...

Страница 13: ...0 the probe is broken and has to be replaced Press the Check key If the meter is calibrated Probe is OK message is displayed If cleaning is needed Clean Probe and Calibrate message is displayed 1 imm...

Страница 14: ...I7009L pH9 18 buffer solution 500 mL bottle HI7010L pH10 01 buffer solution 500 mL bottle CONDUCTIVITY TDS CALIBRATION SOLUTIONS HI7031L 1413 S cm solution 500 mL bottle HI7442L 1500 ppm mg L solution...

Страница 15: ...TIONS FOR USERS Before using this product make sure it is entirely suitable for your specific application and for the environment in which it is used Any variation introduced by the user to the suppli...

Страница 16: ...ark 584 Park East Drive Woonsocket RI 02895 USA www hannainst com Local Office Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA Phone 800 426 6287 Fax 401 765...

Отзывы: