19
• Turn the meter on by pressing
ON/OFF
.
• When the beeper sounds briefly and the LCD displays dashes, the meter is ready.
• Place the
CAL Check
TM
Standard
HI96750‑11
Cuvette A into the cuvette holder and ensure that
the notch on the cap is positioned securely into the groove.
• Press
ZERO/CFM
and the lamp, cuvette and detector icons will appear on the display, depending
on the measurement phase.
• After a few seconds, the display will show
“-0.0-”
. The meter is now zeroed and ready for
validation.
• Remove the cuvette.
• Place the
CAL Check
TM
Standard
HI96750‑11
Cuvette B into the cuvette holder and ensure that
the notch on the cap is positioned securely into the groove.
Use the validation procedure to ensure that the instrument is properly calibrated.
Warning
: Do not validate the instrument with any standard solutions other than the Hanna
Instruments
CAL Check
TM
Standards, otherwise erroneous results will be obtained.
Note: The validation is performed only for the selected parameter. For full validation of the
instrument, the following procedure must be performed for each parameter.
VALID
ATION PROCEDURE