Hanna Instruments HI 96706C Скачать руководство пользователя страница 18

1 8

• After a few seconds the display will show

“-0.0-”

. The meter is now zeroed and ready

for calibration. The blinking 

“READ”

 asks for

reading calibration standard.

• Remove the cuvet.

• Place the 

CAL CHECK

TM

 Standard 

HI 95706-11

Cuvet B into the cuvet holder and ensure that
the notch on the cap is positioned securely into
the groove.

• Press 

READ/ /TIMER

 and the lamp, cuvet

and detector icons will appear on the display,
depending on the measurement phase.

• After measurement the instrument will show

for three seconds the Cal Check Standard value.

Note:

 If the display shows 

“STD HIGH”

, the

standard value was too high. If the display
shows

“STD LOW”

, the standard value was

too low. Verify that both 

CAL CHECK

TM

Standards 

HI 95706-11

 Cuvets, A and B are

free from fingerprints or dirt and that they are
inserted correctly.

   Then the date of the last calibration (e.g.:

”01.08.2005”

) appears on the display, or

“01.01.2005”

 if the factory calibration was

selected before. In both cases the year number
is blinking, ready for date input.

Содержание HI 96706C

Страница 1: ...1 w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives Instruction Manual HI 96706C Phosphorus ISM...

Страница 2: ...REPLACEMENT 22 ACCESSORIES 22 WARRANTY 23 Dear Customer Thank you for choosing a Hanna product This manual will provide you with the necessary information for the correct use of the instrument Please...

Страница 3: ...your Dealer Each HI 96706 Ion Selective Meter is supplied complete with Two Sample Cuvets and Caps 9V Battery Scissors Tissue for wiping cuvets Instrument quality certificate Instruction Manual Rigid...

Страница 4: ...his makes the instruments fulfill field applications Display messages aid the user in routine operation The meter has an auto shut off feature that will turn off the instrument after 10 minutes of non...

Страница 5: ...ur of non use in calibration mode with last reading reminder Dimensions 192 x 102 x 67 mm 7 6 x 4 x 2 6 Weight 290 g 10 oz REQUIRED REAGENTS Code Description Quantity test HI 93706A 0 Phosphorus Reage...

Страница 6: ...t at wavelength c molar concentration of the substance d optical path through the substance Therefore the concentration c can be calculated from the absorbance of the substance as the other factors ar...

Страница 7: ...incoming value into the desired measuring unit and to display it on the LCD The measurement process is carried out in two phases first the meter is zeroed and then the actual measurement is performed...

Страница 8: ...ime OPERATING MODES Measurement mode default operation mode enables both validation and measurement Calibration mode may be entered by keeping CAL CHECK pressed for three seconds the CAL tag appears i...

Страница 9: ...rm a measurement Please check the preparation of the zero cuvet Light Low There is not enough light to perform a measurement Please check the preparation of the zero cuvet No Light The instrument cann...

Страница 10: ...n over range condition The concentration of the sample is beyond the programmed range dilute the sample and re run the test c during calibration procedure Standard Low The standard reading is less tha...

Страница 11: ...is important that the sample does not contain any debris This would corrupt the reading Each time the cuvet is used the cap must be tightened to the same degree Whenever the cuvet is placed into the...

Страница 12: ...gently tapping the cuvet Do not let the reacted sample stand too long after reagent is added or accuracy will be lost It is possible to take multiple readings in a row but it is recommended to take a...

Страница 13: ...bidity or color the measurement takes place in two phases First the meter is zeroed using the unreacted sample After the reagents are added the reacted sample is measured When the beeper sounds briefl...

Страница 14: ...A 0 Phosphorus Reagent A to the cuvet Add the content of one packet of HI 93706B 0 Phosphorus Reagent B to the cuvet Replace the cap and shake gently until dissolution is complete Replace the cuvet in...

Страница 15: ...100 mg L as Fe2 Magnesium Mg2 above 40 000 mg L as CaCO3 Sulfide S2 VALIDATION PROCEDURE Use the validation procedure to ensure that the instrument is properly calibrated Warning Do not validate the...

Страница 16: ...n the display depending on the measurement phase At the end of the measurement the display will show the validation standard value The reading should be within specifications as reported in the CAL CH...

Страница 17: ...OFF When the beeper sounds briefly and the LCD displays dashes the meter is ready Press and hold CAL CHECK for three seconds to enter calibration mode The display will show CAL during calibration proc...

Страница 18: ...isplay depending on the measurement phase After measurement the instrument will show for three seconds the Cal Check Standard value Note If the display shows STD HIGH the standard value was too high I...

Страница 19: ...atically increased When the correct year has been set press ZERO CFM or READ TIMER to confirm Now the display will show the month blinking Press GLP to edit the desired month 01 12 If the key is kept...

Страница 20: ...on the secondary display LAST CALIBRATION DATE If no calibration was performed the factory calibration message F CAL will appear on the main display and the instrument returns to measurement mode aft...

Страница 21: ...a battery indicator with three levels as follows 3 lines for 100 capacity 2 lines for 66 capacity 1 line for 33 capacity Battery icon blinking if the capacity is under 10 If the battery is empty and a...

Страница 22: ...tery cover and turn it clockwise to close To replace the instrument s battery follow the steps ACCESORIES REAGENT SET HI93706 01 Reagentsfor100tests HI93706 03 Reagentsfor300tests OTHER ACCESORIES HI9...

Страница 23: ...ntained according to the instructions This warranty is limited to repair or replacement free of charge Damages due to accident misuse tampering or lack of prescribed maintenance are not covered If ser...

Страница 24: ...2 4 MAN96706 03 06 w w w h a n n a i n s t c o m...

Отзывы: