background image

11

ERROR  MESSAGES

a) on zero reading

This indicates that the zeroing procedure failed due to a low signal-
to-noise ratio. In this case press ZERO again.

“No Light”: the instrument cannot adjust the light level. Please check

that the sample does not contain any debris.

“Light Low”: there is not enough light to perform a measurement.
Please check the preparation of the zero cuvet.

“Light High”: there is too much light to perform a measurement.
Please check the preparation of the zero cuvet.

b) on sample reading

There is too much light for the sample measurement. Please check
if the right sample cuvet is inserted.

“Inverted”: the sample and the zero cuvet are inverted.

“Zero”: a zero reading was not taken. Follow the instruction in
the 

measurement procedure

 for zeroing the meter.

Under range. A blinking "0" for 771 or “0.00” for 701 indicates
that the sample absorbs less light than the zero reference. Check
the procedure and make sure you use the same cuvet for reference
(zero) and measurement.

A flashing value of the maximum concentration (i.e. “500” for
chlorine Ultra High Range 771 or “5.00” for chlorine Low Range
701) indicates an over range condition. The concentration of the
sample is beyond the programmed range: dilute the sample and
re-run the test.

A flashing value lower than the maximum concentration (i.e.
“200” for chlorine Ultra High Range 771 or “2.00” for chlorine

Low Range 701) indicates a  low signal-to-noise ratio condition.
In this case accuracy of the result is not guaranteed. Repeat the

measurement procedure

.

14

MEASUREMENT PROCEDURE

MEASUREMENT PROCEDURE FOR 771

• Select the parameter code “771” for Chlorine

ULTRA HIGH RANGE by holding ZERO/SET
(see page 13).

• When the LCD displays “771”, it is ready.

• Fill the cuvet with 10 mL of unreacted

sample, up to the mark, and replace the
cap.

• Place the cuvet into the holder and ensure

that the notch on the cap is positioned
securely into the groove.

• Press ZERO/SET and “SIP” will blink on the

display.

• After a few seconds the display will show

“-0.0-”. The meter is now zeroed and ready for
measurement.

• Remove the cuvet from the instrument. Remove

the cap.

• Add the content of one packet of

HI 95771A-0 Chlorine Ultra HR reagent A.

• Add the content of one packet of

HI 95771B-0 Chlorine Ultra HR reagent B.
Replace the cap and shake gently for 20

seconds.

Содержание HI 95771C

Страница 1: ...w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives Instruction Manual HI 95771C Chlorine Ultra High Range ISM MAN95771CR3 06 06 w w w h a n n a i n s t c o m...

Страница 2: ...choosing a Hanna product This manual will provide you with the necessary information for the correct use of the instrument Please read it carefully before using the meter If you need additional techn...

Страница 3: ...o avoid damages or burns do not put the instrument in microwave oven For yours and the instrument safety do not use or store the instrument in hazardous environments HI 95771 is warranted for two year...

Страница 4: ...m range and in the 0 to 500 mg L ppm range The methods are an adaptation of Standard Methods for the Examination of Water and Wastewater 20th edition 4500 Cl The reagents are in powder form and are su...

Страница 5: ...Press READ TIMER and SIP will blink during measurement The instrument will show for three seconds the CAL CHECK standard value i e 100 for 771 chlorine ULTRA HIGH RANGE or 1 00 for 701 chlorine LOW R...

Страница 6: ...through the substance Therefore the concentration c can be calculated from the absorbance of the substance as the other factors are known 19 Note It is possible to interrupt calibration procedure at a...

Страница 7: ...thus requires particular attention It is important that both the measurement and the calibration zeroing cuvets are optically identical to provide the same measurement conditions Whenever possible use...

Страница 8: ...andards otherwiseerroneousresultswillbeobtained For accurate Validation and Calibration results please perform test at room temperature 18 to 25 C 64 5 to 77 0 F VALIDATION PROCEDURE Hold READ TIMER f...

Страница 9: ...dealer or the nearest Hanna Customer Service Center GUIDE TO DISPLAY CODES 16 MEASUREMENT PROCEDURE Select the parameter code 701 for Chlorine LOW RANGE by holding ZERO SET see page 13 When the LCD d...

Страница 10: ...econd at the end of the calibration procedure to indicate that the calibration data has been stored Error the concentration of the calibration solution used is not correct Repeat the calibration proce...

Страница 11: ...ne Ultra High Range 771 or 5 00 for chlorine Low Range 701 indicates an over range condition The concentration of the sample is beyond the programmed range dilute the sample and re run the test A flas...

Страница 12: ...recommended to take a new zero reading for each sample and to use the same cuvet for zeroing and measurement After the reading it is important to discard immediately the sample otherwise the glass mig...

Отзывы: