The flashing digits indicate that:

1) the month can be set
2) the day can be set

“Storage”: this prompt appears for 1 second at the end of the

calibration procedure

, to indicate that the calibration data has

been stored.

“Error”: the concentration of the calibration solution used is not
correct. Repeat the 

calibration procedure

 with the right standard

solution, and verify it is not expired. If the calibration procedure fails
again, contact your dealer or the nearest Hanna Customer Service
Center.

ERROR  MESSAGES

a) on zero reading

This indicates that the zeroing procedure failed due to a low signal-
to-noise ratio. In this case press ZERO again.

“No Light”: the instrument cannot adjust the light level. Please check
that the sample does not contain any debris.

“Light Low”: there is not enough light to perform a measurement.
Please check the preparation of the zero cuvet.

“Light High”: there is too much light to perform a measurement.
Please check the preparation of the zero cuvet.

b) on sample reading

There is too much light for the sample measurement. Please check
if the right sample cuvet is inserted.

“Inverted”: the sample and the zero cuvet are inverted.

“Zero”: a zero reading was not taken. Follow the instruction in
the 

measurement procedure

 for zeroing the meter.

11

Under range. A blinking "0.00" indicates that the sample absorbs
less light than the zero reference. Check the procedure and make
sure you use the same cuvet for reference (zero) and measurement.

A  flashing value of the maximum concentration indicates an
over range condition. The concentration of the sample is beyond
the programmed range: dilute the sample and re-run the test.

A flashing value lower than the maximum concentration indicates
a  low signal-to-noise ratio condition. In this case accuracy of the
result is not guaranteed. Repeat the 

measurement procedure

.

The instructions listed below should be carefully followed during testing to ensure best accuracy.

• Color or suspended matter in large amounts may

cause interference, therefore these should be
removed by treatment with active carbon and by
prior filtration.

• For a correct filling of the cuvet: the liquid in

the cuvet forms a convexity on the top; the
bottom of this convexity must be at the same
level of the 10 mL mark.

• Proper use of the dropper:

(a) to get good reproducible results, tap the dropper on the table for several times and wipe

the outside of the dropper tip with a cloth.

(b) always keep the dropper bottle in a vertical position while dosing the reagent.

GENERAL TIPS FOR AN ACCURATE MEASUREMENT

(a)

(b)

Содержание HI 95715

Страница 1: ...w w w h a n n a c a n c o m This Instrument is in Compliance with the CE Directives Instruction Manual HI 95715 Ammonia Medium Range ISM MAN95715R1 09 03 w w w h a n n a c a n c o m...

Страница 2: ...ry steps to correct interferences Any variation introduced by the user to the supplied equipment may degrade the instruments EMC performance To avoid damages or burns do not put the instrument in micr...

Страница 3: ...s 1 set HI 710009 Blue rubber boot HI 710010 Orange rubber boot HI 721310 9V battery 10 pcs HI 731318 Tissue for wiping cuvets 4 pcs HI 731331 Glass cuvets 4 pcs HI 731335 Caps for cuvets 4 pcs HI 937...

Страница 4: ...ater samples in the 0 00 to 9 99 mg L ppm range The method is an adaptation from ASTM Manual of Water and Environmental Technology D1426 93 Nessler Method The reagents are in liquid form and are suppl...

Страница 5: ...ry using a standard solution of 6 00 mg L ammonia nitrogen NH3 N and a representative lot of reagents an operator obtained with a single instrument a standard deviation of 0 12 mg L PRECISION AND ACCU...

Страница 6: ...a narrow band interference filter to guarantee both high performance and reliable results HI 95715 block diagram optical layout PRINCIPLE OF OPERATION 15 VALIDATION PROCEDURE Turn the meter on by pres...

Страница 7: ...ce due to unwanted reflection and absorption of light It is recommended not to touch the cuvet walls with hands Furthermore in order to maintain the same conditions during the zeroing and the measurin...

Страница 8: ...Day icons appear above the number of month and day when a date is displayed 4 the blinking Battery icon appears when the battery voltage is getting low 13 MEASUREMENT PROCEDURE Turn the meter on by p...

Страница 9: ...n is selected GUIDE TO DISPLAY CODES 12 It is important that the sample does not contain any debris This would corrupt the reading In order to avoid reagent leaking and to obtain more accurate measure...

Страница 10: ...s not taken Follow the instruction in the measurement procedure for zeroing the meter 11 Under range A blinking 0 00 indicates that the sample absorbs less light than the zero reference Check the proc...