This prompt appears for 1 second each time the instrument is

turned on.

This indicates that the meter is in a ready state and zeroing can
be performed.

Sampling in Progress. This 

flashing prompt

 appears each time the

meter is performing a measurement.

This indicates that the meter is in a zeroed state and measurement
can be performed.

Light over range: the cuvet is not inserted correctly and an excess

ambient light is reaching the detector. If the cover is properly
installed, then contact your dealer or the nearest Hanna Customer
Service Center.

The blinking "BAT" indicates that the battery voltage is getting low

and the battery needs to be replaced.

This indicates that the battery is dead and must be replaced. Once

this indication is displayed, the meter will lock up. Change the
battery and restart.

“Configuration”: this indicates that the meter has lost its
configuration. Contact your dealer or the nearest Hanna Customer

Service Center.

CALIBRATION MODE  MESSAGES

The date of the last calibration appears on the LCD each time the

calibration mode

 is entered. If calibration is performed for the first

time, “F.CAL” appears.

This indicates that the factory calibration is selected.

GUIDE TO DISPLAY CODES

12

• Proper use of the dropper:

(a) to get good reproducible results, tap the dropper on the table for several times and wipe

the outside of the dropper tip with a cloth.

(b) always keep the dropper bottle in a vertical position while dosing the reagent.

• Proper use of the powder reagent packet:

(a) use scissors to open the powder packet;
(b) push the edges of the packet to form a spout;
(c) pour  out the content of the packet.

• Shaking the cuvet can generate bubbles in the sample, causing higher readings. To obtain

accurate measurements, remove such bubbles by swirling or by gently tapping the vial.

• Do not let the reacted sample stand too long after reagent is added, or accuracy will be lost.

• It is possible to take multiple readings in a row, but it is recommended to take a new zero

reading for each sample and to use the same cuvet for zeroing and measurement.

• After the reading it is important to discard immediately the sample, otherwise the glass

might become permanently stained.

• All the reaction times reported in this manual are referred to 20°C (68°F). As a general rule

of thumb, they should be doubled at 10°C (50°F) and halved at 30°C (86°F).

• In order to maximize accuracy, prior to a measurement follow the 

validation procedure

 to

be sure that the instrument is properly calibrated. If necessary, calibrate the instrument.

(a)

(b)

(a)

(b)

(c)

Содержание HI 95706C

Страница 1: ...1 Instruction Manual HI 95706C Phosphorus ISM w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives 20 MAN95706R2 06 06 w w w h a n n a i n s t c o m...

Страница 2: ...take all necessary steps to correct interferences Any variation introduced by the user to the supplied equipment may degrade the instruments EMC performance To avoid damages or burns do not put the i...

Страница 3: ...ACCESSORIES HI 95706 11 CAL CHECK Standard Cuvets 1 set HI 710009 Blue rubber boot HI 710010 Orange rubber boot HI 721310 9V battery 10 pcs HI 731318 Tissue for wiping cuvets 4 pcs HI 731331 Glass cuv...

Страница 4: ...asures the phosphorus P content in water samples in the 0 0 to 15 0 mg L ppm range The method is an adaptation from Standard Method for the Examination of Water and Wastewater The reagents are in powd...

Страница 5: ...tandardsolutionof6 00 mg L phosphorus and a representative lot of reagents an operator obtained with a single instrumentastandarddeviationof0 2mg L SPECIFICATIONS PRECISION AND ACCURACY 16 CALIBRATION...

Страница 6: ...band interference filters to guarantee both high performance and reliable results HI 95 series block diagram optical layout PRINCIPLE OF OPERATION 15 Turn the meter on by pressing ON OFF When the LCD...

Страница 7: ...ever possible use the same cuvet for both It is necessary that the surface of the cuvet is clean and not scratched This to avoid measurement interference due to unwanted reflection and absorption of l...

Страница 8: ...ation mode 3 the Month and Day icons appear above the number of month and day when a date is displayed 4 the blinking Battery icon appears when the battery voltage is getting low 13 MEASUREMENT PROCED...

Страница 9: ...ration is selected GUIDE TO DISPLAY CODES 12 Proper use of the dropper a to get good reproducible results tap the dropper on the table for several times and wipe the outside of the dropper tip with a...

Страница 10: ...ure for zeroing the meter 11 Under range A blinking 0 00 indicates that the sample absorbs less light than the zero reference Check the procedure and make sure you use the same cuvet for reference zer...

Отзывы: