Hanna Instruments HI 935002 Скачать руководство пользователя страница 5

9

8

HI935009

HI935005 / 5005N

OPERATIONAL GUIDE

To switch the instrument ON, press the ON/OFF key. 
The thermometer will carry out a self diagnostic test routine, the LCD will 
show all segments for a few seconds (or as long as ON/OFF is held), 
followed by the percentage indication of the remaining battery life.
The thermometer then enters normal measurement mode. 

If a temperature probe is plugged in, the meter displays the measured 
temperature. 
If no probe is plugged in, or if the reading is over-range, the display 
shows flashing dashes. 
If a measurement is slightly over the range of 
the meter specifications, the display will flash 
the closest full-scale value.
To switch the meter OFF, press the ON/OFF key.

°C/°F SELECTION

Measurements can be displayed in either degrees Celsius or Fahrenheit. 
The meter is factory set to °C scale; to change the scale, press °C/°F 
key or (ALT+) °C/°F keys, or set the switch located in the battery 
compartment (

HI935002 

only, see figure on page 10). 

HOLD and MR FUNCTIONS

The HOLD function is activated by pressing the HOLD key. 
The measured temperature is held on the display until HOLD is pressed 
again. The "HOLD" tag blinks on the display while in HOLD mode.

Note

: Although the display is frozen, internally the meter continues 

measuring and updating Hi/Lo and relative values.

In 

HI935009

 only, the held value is also stored in non-volatile memory 

and can be recalled by pressing (ALT+) MR key. While the MR key 
is held, the "HOLD" tag lights up and the meter displays the stored 
value; when the MR key is released, after 1 second the meter returns 
to normal measurement mode.

HIGH/LOW TEMPERATURES

The maximum and minimum temperatures 
are continuously monitored and displayed 
on the lower portion of the LCD. 

Note

: When the reading goes over-range or the probe is removed, 

the Hi and Lo values display dashes until cleared. 

CLEAR FUNCTION

Upon pressing the CLR key, the High/Low values may be cleared at any 
time during measurement and the current reading is assigned to the 
highest and lowest temperature values for the displayed channel only.

AUTO SHUT-OFF (HI935005 and HI935009)

To save battery life, the 

HI935005

 and 

HI935009

 models are provided 

with an auto-off feature, which switches the instrument off after a 
certain period of nonuse.
The delay time can be changed by the user (8 minutes, 60 minutes, 
or disabled). 
To access this feature, the 

HI935005

 is provided with an internal button 

located in the battery compartment (see figure on page 10). Unscrew 
the 4 screws on the back of the meter and select the desired auto-off 
time period by pressing the button and viewing the value on the LCD. 
Re-attach the back making sure that the gasket is in place and tighten 
the 4 screws to ensure a watertight seal. 

Содержание HI 935002

Страница 1: ...Instruction Manual HI935005 HI935005N HI935002 HI935009 Portable K Thermocouple Thermometers www hannainst com ...

Страница 2: ...which have been designed using the latest microprocessor technology to provide reliable and accurate measurements in a wide temperature range Standard features include K type probes waterproof casing dual level LCD low battery detection long battery life and two year warranty HI935005 has been designed for laboratory and field use while HI935005N is suitable for high accuracy meter and probe can b...

Страница 3: ...ear the HI and LO values BL HI935005N only toggle the backlight ON and OFF CAL HI935005N only press and hold for 5 seconds to enter the Calibration mode with reading within 3 C range press to abort the Calibration mode Range may be limited by probe SPECIFICATIONS Range 50 0 to 199 9 C 200 to 1350 C 58 0 to 399 9 F 400 to 2462 F Resolution 0 1 C up to 199 9 C 1 C outside 0 1 F up to 399 9 F 1 F out...

Страница 4: ...ement DISP select Normal Relative or T1 T2 measuring mode ALT HI935009 only enable the second function keys the ALT tag turns on to indicate that the second functions are enabled Note The ALT key can be released before pressing the second function key for 1 hand operation ALT CAL press and hold for about 5 seconds to enter the Calibration mode with reading within 3 C range ALT AOFF set auto power ...

Страница 5: ...d updating Hi Lo and relative values In HI935009 only the held value is also stored in non volatile memory and can be recalled by pressing ALT MR key While the MR key is held the HOLD tag lights up and the meter displays the stored value when the MR key is released after 1 second the meter returns to normal measurement mode HIGH LOW TEMPERATURES The maximum and minimum temperatures are continuousl...

Страница 6: ...e corresponding tags will light up to inform the user To select the desired channel use the T1 T2 key The 2 channel models are also provided with the DISP function which allows the user to select the information to be displayed While in T1 or T2 pressing DISP will switch the display between standard and relative measurement modes In the standard mode the main body of the LCD shows the current temp...

Страница 7: ...ce made from distilled water Immerse the temperature probe in the center of the ice bath taking care not to touch the ice with the probe tip Ensure that the meter is measuring a temperature within 3 C To enter the Calibration mode press and hold the CAL button for about 5 seconds The CAL tag turns on to indicate that the Calibration mode has been entered Note If the measurement is outside the 3 C ...

Страница 8: ...unction with the HI766HD probe handle HI766PA Roller surface probe max 320 C 600 F HI766PB Surface probe max 650 C 1200 F HI766PC Penetration probe max 900 C 1650 F HI766PD Air probe max 300 C 570 F HI766PE1 General purpose probe max 900 C 1650 F HI766PE2 General purpose probe max 900 C 1650 F grill surface probe HI766B4 Grill surface probe with 70 cm 27 6 cable protect ed with stainless steel jac...

Страница 9: ...truments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA www hannainst com Local Office Hanna Instruments Inc Highland Industrial Park 584 Park East Drive Woonsocket RI 02895 USA Phone 800 426 6287 Fax 401 765 7575 e mail tech hannainst com ...

Отзывы: