Hanna Instruments HI 931101 Скачать руководство пользователя страница 9

17

16

  • To quit the calibration and save

the first point, press the CAL
key.

For better accuracy however, it is recommended
that a two-point calibration is performed.

TWO POINT CALIBRATION

    • After the first point is confirmed,

immerse the FC 300B electrode
into the second solution:
HI 7085 at 0.3 g/L NaCl or
HI 7081 at 30 g/L NaCl.

Note: The electrode should be

immersed approximately
2.5 cm (1") into the so-
lution.

 • Select the second point value

by pressing the BUF key. Make
sure that the selected value cor-
responds to the one being used
for the second calibration point.

 • If the reading is not close to

the selected value, the "BUF"
symbol will blink.

Otherwise, if the reading is
close to the selected value, the
meter will recognize the stan-
dard solution. Only the "

"

symbol will blink, to indicate
that the reading is stabilizing.

Once stabilized "CON" symbol
will appear on the display.

For best results, wait for a
couple of minutes and press
the CFM key.

  • The two point calibration is now

completed and the meter re-
turns to the normal mode.

CAL

BUF

SINGLE POINT CALIBRATION (OFFSET)

 • Dip the FC 300B sodium elec-

trode into the HI 7083 (3.00 g/L
NaCl) solution.

Note: Immerse the electrode

approx. 2.5 cm (1") into
the solution.

Press the CAL key. The "CAL" and "BUF" indi-
cators will be displayed with the "BUF" and the
"

"  symbols blinking.

 • As soon as the standard solu-

tion is recognized, only the
"

"  symbol will continue

blinking to indicate that the elec-
trode is stabilizing.

The value of the standard solu-
tion being used will be displayed
on the primary LCD.

  • As soon as the electrode is sta-

bilized, "CON" symbol lights up.

For best results, wait for a couple
of minutes before pressing the
CFM key.

  • HI  931100 will then display "30.0

gm".

CAL

gm

°C

NaCl

BUF

CAL

CFM

CFM

Содержание HI 931101

Страница 1: ...1 Instruction Manual HI 931100 HI 931101 Waterproof Instruments for Salinity and Sodium Content Measurements w w w h a n n a i n s t c o m...

Страница 2: ...HI 931101 employs the FC 300B combination sodium electrode to provide sodium readings from 15 0 mg L to 60 g L The calibration is automatic at 2 points the first is at 2 3 g L while the second can be...

Страница 3: ...to 300 g L NaCl 0 0 to 80 0 C Resolution 0 001 0 01 0 1 1 g L NaCl 0 1 C Accuracy 5 of reading NaCl 20 C 0 5 C Typical EMC Deviation 2 FS NaCl 0 5 C Calibration Automatic 1 or 2 point at 3 00 g L HI...

Страница 4: ...Compensation 25 C 77 F Electrode FC300B sodium electrode with BNC an 1 m 3 3 cable not included Temperature Probe HI 7662 not included Input Impedance 1012 Ohm Battery Type Life 4 x 1 5V AA approx 200...

Страница 5: ...is observed for at least a couple of minutes During measurement if possible the sample should be gently stirred 100 rpm OPERATIONAL GUIDE INITIAL PREPARATION Each meter is supplied com plete with four...

Страница 6: ...to pNa Note that pNa log aNa where aNa is the activity of ion Na Press the MODE key again to return to g L mea surements TEMPERATURE MEASUREMENTS Connect the temperature probe to the meter and turn on...

Страница 7: ...e performed at 1 offset only or 2 points For best accuracy a 2 point procedure is recom mended PREPARATION Pour small quantities of HI7080 2 3 g L Na solution and HI 7087 0 23 g L Na or HI 7086 23 g L...

Страница 8: ...eter will recognize the stan dard solution and only the symbol will blink to indicate that the reading is stabilizing Once stabilized CON symbol will appear on the display BUF CAL NaCl CALIBRATION HI...

Страница 9: ...ll blink to indicate that the reading is stabilizing Once stabilized CON symbol will appear on the display For best results wait for a couple of minutes and press the CFM key The two point calibration...

Страница 10: ...wer than 5 g L it is recommend to use the HI 7090 ISA solution In such case calibration should be per formed with standard solution containing 2 of ISA ISA Ionic Strength Adjuster 50 of NH4 Cl 4M 50 o...

Страница 11: ...e calibration mode press the CAL and ON OFF keys simultaneously Immerse the temperature probe into the 0 C 32 F bath together with the reference thermom eter When the reading stabilizes press the CAL...

Страница 12: ...elow the filling hole add HI 8093 1M KCl AgCl electrolyte solution Shift downwards the rub ber sleeve during mea surements and calibra tion SODIUM MEASUREMENT Rinse the sodium electrode tip with some...

Страница 13: ...tes rinse thor oughly in a solution with 2 of ISA and then follow the Cleaning Procedure The sensitive bulb should not touch the bottom of the beaker The Hanna HI 76405 electrode holder can be used fo...

Страница 14: ...olution 230 mL FDA bottle HI 7083L 3 0 g L NaCl solution 500 mL bottle HI 8083L 3 0 g L NaCl solution 500 mL FDA bottle HI 7085M 0 3 g L NaCl solution 230 mL bottle HI 8085M 0 3 g L NaCl solution 230...

Страница 15: ...rring capacity speed range from 100 to max 1000 rpm HI 710031 Rugged carrying case HI 7662 Stainlesssteeltemperatureprobewith 1 m 3 3 cable HI 76405 Electrodeholder All Hanna Instruments meters are wa...

Страница 16: ...to radio and TV equipments requiring the operator to take all necessary steps to correct interferences Any variation introduced by the user to the supplied equipment may degrade the instruments EMC pe...

Страница 17: ...258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax 011 615 8582 Taiwan Tel 886 2 2739...

Отзывы: