Hanna Instruments HI 931100 Скачать руководство пользователя страница 9

17

17

 • If the reading is not close

to the selected buffer, the
"BUF" prompt will continue
to blink.

If on the other hand, the
reading is close to the se-
lected buffer, the meter
will recognize the standard
solution. The BUF symbol
will stay still on the dis-
play, while the "

" sym-

bol will blink, indicating that
the meter is stabilizing its
reading.

 • Once stabilized "CON"

symbol will appear on
the display.

Wait for a couple of min-
utes and press the CFM
key.

 • The meter will return to its normal mode

and the two point calibration process is
completed.

CFM

mannacl_r1.P65

18

18

NaCl CALIBRATION (HI931100/HI931102)

The calibration is performed by immersing
the 

FC 300B

 sodium-sensitive electrode

 

in 2

different standard NaCl solutions to achieve
accurate 2 point calibration.
It is also possible to calibrate just the offset
(3.00 g/L NaCl) following the single point cali-
bration procedure described below.
For best accuracy, a two-point calibration is
recommended (see below).

PREPARATION

  •

For HI931100 and HI931102:

Pour small quantities of

HI 7083 

or 

HI8083 

(3.00 g/L

NaCl) solution and 

HI7085

or 

HI 8085

 (0.30 g/L) or

HI 7081 

or 

HI 8081 

(30 g/L)

into two clean beakers.
To get accurate readings use HI7083/HI8083
and HI 7085/HI8085 if you are going to mea-
sure samples with low sodium chloride con-
tent (below 3 g/L) or HI7083/HI8083 and
HI7081/HI8081 if you are going to measure
samples with high sodium chloride content
(over 3 g/L).

  • For accurate calibra-

tion, use two bea-
kers for each buffer
calibration, the first
one for rinsing the electrode, the second
one for calibration. This way, contamina-
tion of the buffer is mini-
mized.  Whenever possible
use plastic beakers to mini-
mize any EMC interferences.

 • Turn the meter on by

pressing the ON/OFF key.

  • Remove the protective cap

of the electrode and rinse

HI 7083

HI 7085

H

7

0

8

3

ON/OFF

ON/OFF

Содержание HI 931100

Страница 1: ...ise idea of its versatility in a wide range of applications These instruments are in compliance with directives TABLEOFCONTENTS TABLEOFCONTENTS TABLEOFCONTENTS TABLEOFCONTENTS TABLEOFCONTENTS Prelimin...

Страница 2: ...cuitry has a built in protection against elec tromagnetic interference All meters come equipped with a large and easy to read LCD which shows sodium so dium chloride and temperature values simul taneo...

Страница 3: ...and 0 30 g L HI 7085 or 30 0 g L HI 7081 fixed at 25 C 77 F FC 300B glass combination sodium sensitive electrode not included HI 7662 temperature probe not included 4 x 1 5V AA size 200 hours of cont...

Страница 4: ...50 C 32 to 122 F 100 RH 196 x 80 x 60 mm 7 7 x 3 1 x 2 4 450 g 15 8 lb Measurement pNa Na C Resolution pNa Na C Accuracy pNa 20 C 68 F Na C Typical EMC pNa Deviation Na C Calibration Temperature Compe...

Страница 5: ...PREPARATION INITIAL PREPARATION INITIAL PREPARATION INITIAL PREPARATION INITIAL PREPARATION Each meter is supplied complete with 4x1 5V AA batteries Remove the back cover unwrap the batteries and inst...

Страница 6: ...For greater accuracy per form the calibration of the meter with stan dard solution at a temperature as close as possible to the temperature of the sample 1 C 2 F TAKING SODIUM MEASUREMENTS TAKING SOD...

Страница 7: ...ion is performed by immersing the FC300B sodium sensitive electrode in 2 different standard Na solutions to achieve ac curate 2 point calibration It is also possible to calibrate just the offset 2 3 g...

Страница 8: ...CON symbol will appear on the display 2 3 g L Na CAL mannacl_r1 P65 16 16 For best results wait for a couple of minutes before pressing the CFM key HI 931101 will display 0 230 gm If the CAL key is pr...

Страница 9: ...et 3 00 g L NaCl following the single point cali bration procedure described below For best accuracy a two point calibration is recommended see below PREPARATION For HI931100 and HI931102 Pour small q...

Страница 10: ...rimary LCD As soon as the electrode is stabilized CON sym bol will appear on the dis play For best results wait for a couple of minutes before CAL CAL 3 g L NaCl mannacl_r1 P65 20 20 pressing the CFM...

Страница 11: ...display For best results wait for a couple of minutes and press the CFM key The meter will return to its normal mode and the two point calibration process is completed BUF BUF CFM CFM mannacl_r1 P65...

Страница 12: ...y the electrode to stabilize The next measurements in cheese or meat samples will have a shorter response time less than 2 min utes Do not rinse the sodium elec trode between measurements since con ta...

Страница 13: ...ture probe HI7662 to the RCA socket on the top of the instru ment Turn the instrument off Press the CAL and the ON OFF keys simulta neously 0 C 32 F C 50 C 122 F C gm C NaCl CAL CAL ON OFF ON OFF mann...

Страница 14: ...he protective cap DO NOT BE ALARMED IF ANY SALT DEPOSITS ARE PRESENT This is normal with sodium electrodes and they will disap pear when rinsed with water During transport tiny bubbles of air may have...

Страница 15: ...t samples 1 or 2 times per week store the electrode in HI7081 or HI8081 30 0 g L NaCl if used for NaCl measurements in conjunction with HI931100 or HI931102 Store the electrode in HI7080 or HI8080 2 3...

Страница 16: ...LOT DRY THE ELECTRODE TROUBLESHOOTING Evaluate your electrode performance based on the following Noise Readings fluctuate up and down could be due to Clogged Dirty Junction Refer to the Cleaning Proce...

Страница 17: ...2 3 g L Na solution 230 mL HI 7080L 2 3 g L Na solution 460 mL HI 7086M 23 g L Na solution 230 mL HI 7086L 23 g L Na solution 460 mL HI 7087M 0 230 g L Na solution 230 mL HI 7087L 0 230 g L Na solutio...

Страница 18: ...norganic Cleaning Sol 460 mL HI7077M Oil Fat Cleaning Sol 230 mL HI7077L Oil Fat Cleaning Sol 460 mL mannacl_r1 P65 36 36 ELECTRODE CLEANING SOLUTIONS IN FDA APPROVED BOTTLES HI8061M General Cleaning...

Страница 19: ...rom whom you purchased the instrument If under warranty report the model number date of purchase serial number and the nature of the failure If the repair is not covered by the war ranty you will be n...

Страница 20: ...ace exceed 24VAC or 60VDC To avoid damages or burns do not perform any measurement in microwave ovens DECLARATION OF CONFORMITY We Hanna Instruments Srl V le delle industrie 12 35010 Ronchi di Villafr...

Отзывы: