15
Press
Modify
to access the baud rate screen. Use
ARROW
to scroll through baud rate values and
press
Accept
to confirm the selection or
ESC
to
return to the setup menu without saving the new
value.
Options
:
600
,
1200
,
2400
,
4800
,
9600,
19200, 38400
bps.
This parameter is used for setting the speed of
the serial link between the instrument and a PC.
Both must use the same setting.
Baud rate
Press
Select
to view the firmware version, language
version, mV factory calibration date and time and
temperature factory calibration date and time.
Press
ESC
to return to the
Setup
mode.
Meter information
ELECTRODE PREPARATION
PREPARATION PROCEDURE
Remove the electrode protective cap.
DO NOT BE ALARMED IF ANY SALT DEPOSITS ARE PRESENT. This is normal with electrodes and
they will disappear when rinsed with distilled water.
During transport tiny bubbles of air may have formed inside the glass bulb. The electrode cannot
function properly under these conditions. These bubbles can be removed by "shaking down" the
electrode as you would do with a glass thermometer.
If the bulb is dry, soak the electrode in
HI 70300
Storage Solution for at least one hour.