Add standard screen
Add the corresponding standard. Press
Read
to start the next
step analysis if
“Reading confirm”
option is enabled in the
Setup
menu. If
“Reading confirm”
option is disabled,
the
instrument will start with the next step analysis.
Wrong standard error screen
This screen is displayed if step 2 and 3 of the current reading is
out of ±2000mV range or if the difference between the stable
readings from the current and the previous step is to small
(±3mV) or, if in auto confirm mode, the standard addition is
not detected.
Wrong input error screen
This screen is displayed if the current reading is out of
±2000mV range.
If the difference between the temperature of the sample
confirmed at the first step and the subsequent readings is
greater than 3°C and the temperature compensation feature is
disabled in
Setup
, the temperature warning icon will blink
and the message
“WARNING: Large temperature change!”
will be displayed.
Sample reading confirm screen
(only if this feature is enabled
in setup)
Press
CFM
to memorize the reading.
Pause screen
Press
Resume
to continue the measurement process.
Press
Stir
to start/stop the stirrer.