35
PHOSPHORUS HIGH RANGE
SPECIFICATIONS
Range
0 to 100 mg/L
Resolution
1 mg/L
Accuracy
±5 mg/L ±4% of reading at 25 °C
Typical EMC Dev.
±1 mg/L
Light Source
Tungsten lamp with narrow band interference filter @ 525 nm
Method
Adaptation of the
Standard Methods for the Examination of Water and Wastewater,
18
th
edition, Amino Acid method. The reaction between phosphorus and reagents
causes a blue tint in the sample.
REQUIRED REAGENTS
Code
Description
Quantity
HI 93706
A
-0
Molybdate
10 drops
HI 93706
B
-0
Amino Acid Powder
1 packet
REAGENT SETS
HI 93706-01
Reagents for 100 tests
HI 93706-03
Reagents for 300 tests
For other accessories see page 50.
MEASUREMENT PROCEDURE
Note
: for sample preparation follow the NUTRIENTS SOLUTION (HR) procedure at page 21.
• Select the
Phosphorus HR method using the procedure described
in the
Method Selection section (see page 12).
• Fill the cuvette with 10 mL of unreacted sample (up to the
mark) and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the
Zero
key. The display will show “
-0.0-
” when the meter is zeroed and ready for measurement.
• Remove the cuvette.
• Add 10 drops of HI 937O6A-0 Molybdate reagent.
Phosphorus HR
10 mL