background image

8

9

USING LIQUID AND POWDER REAGENTS

• Proper use of the dropper:

(a) for reproducible results, tap the dropper on the table for several times and wipe  the outside of the

dropper tip with a cloth.

(b) always keep the dropper bottle in a vertical position while dosing the reagent.

• Proper use of the powder reagent packet:

(a) use scissors to open the powder packet;
(b) push the edges of the packet to form a spout;
(c) pour out the content of the packet.

USING CUVETTES

• Proper mixing of the cuvette is done by 

shaking the cuvette

, moving the cuvette up and down. The

movement may be gentle or vigorous. This mixing method is indicated with “shake gently” or “shake
vigorously”, and one of the following icons:

(a)

(b)

The instructions listed below should be carefully followed during testing to ensure most accurate results.
• Color or suspended matter in large amounts may cause interference, they should be removed by

treatment with active carbon and filtration.

• Ensure the cuvette is filled correctly: the liquid in the cuvette forms a convexity on the top; the bottom

of this convexity must be at the same level of the 10 mL mark.

COLLECTING AND MEASURING SAMPLES

• In order to measure exactly 0.5 mL of reagent with the 1 mL syringe:

(a) push the plunger completely into the syringe and insert the tip into the solution.
(b) pull the plunger up until the lower edge of the seal is exactly on the 0.0 mL mark.
(c) take out the syringe and clean the outside of the syringe tip. Be sure that no drops are hanging

on the tip of the syringe, if so eliminate them. Then, keeping the syringe in vertical position above

the cuvette, push the plunger down into the syringe until the lower edge of the seal is exactly on

the 0.5 mL mark. Now the exact amount of 0.5 mL has been added  to the cuvette, even if the

tip still contains some solution.

TIPS FOR AN ACCURATE MEASUREMENT

(c)

(a)

(b)

(c)

(a)

(b)

shake gently

shake vigorously

Содержание HI 83205

Страница 1: ...1 Instruction Manual HI 83205 Multiparameter Bench Photometer for Boilers and Cooling Towers w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives...

Страница 2: ...0 Windows Compatible Software helps users manage all their results Dear Customer Thank you for choosing a Hanna product Please read this instruction manual carefully before using the instrument This m...

Страница 3: ...are optically identical to provide the same measurement conditions Most of methods use the same cuvette for both so it is important that measurements take place in the same optical point The instrumen...

Страница 4: ...elect method menu Press to move up in a menu or a help screen to increment a set value to access second level functions Press to move down in a menu or a help screen to decrement a set value to access...

Страница 5: ...mounts may cause interference they should be removed by treatment with active carbon and filtration Ensure the cuvette is filled correctly the liquid in the cuvette forms a convexity on the top the bo...

Страница 6: ...ods require a reagent blank correction The blank and the sample are prepared exactly in the same way only the blank is deionized water instead of sample A blank cuvette may be used more than once stab...

Страница 7: ...g and deleting You can view and delete the data log by pressing the RCL key Deleting is based on the LIFO last in first out scheme Additionally you can delete the data records all at once CHEMICAL FOR...

Страница 8: ...n enabled a short beep is heard every time a key is pressed A long beep alert sounds when the pressed key is not active or an error condition is detected Press the functional key to enable disable the...

Страница 9: ...05 offers an interactive contextual help mode that assists the user at any time To access help screens pressHELP The instrument will display additional information related to the current screen To rea...

Страница 10: ...into the holder and close the lid Press ZERO key The display will show 0 0 when the meter is zeroed and ready for measurement AMMONIA MEDIUM RANGE Aluminum Add the content of one packet of HI 93712C 0...

Страница 11: ...ide various aliphatic and aromatic amines Ammonia LR SPECIFICATIONS Range 0 00 to 3 00 mg L Resolution 0 01 mg L Accuracy 0 04 mg L 4 of reading Typical EMC 0 01 mg L Deviation Light Source Tungsten l...

Страница 12: ...se the lid Press ZERO key The display will show 0 0 when the meter is zeroed and ready for measurement 10 mL Ammonia LR Remove the cuvette Add 4 drops of HI 93700A 0 First reagent 6 drops for seawater...

Страница 13: ...tte with 10 mL of unreacted sample up to the mark and replace the cap Place the cuvette into the holder and close the lid Free Chlorine FREE CHLORINE 10 mL Remove the cuvette and add the content of on...

Страница 14: ...ey The meter will show 0 0 when the meter is zeroed and ready for measurement Remove the cuvette Powder reagents procedure Add the content of one packet of HI 93701 0 DPD reagent Replace the cap and s...

Страница 15: ...instrument x 3 x 3 x 1 Total Chlorine TOTAL CHLORINE SPECIFICATIONS Range 0 00 to 3 50 mg L Resolution 0 01 mg L Accuracy 0 03 mg L 3 of reading Typical EMC 0 01 mg L Deviation Light Source Tungsten l...

Страница 16: ...rk with the sample Add 0 5 mL of HI 93738A 0 Chlorine Dioxide Reagent to each cylinder 1 2 close them and invert several times to mix 2 25 ml 1 25 ml 1 2 1 2 Press TIMER and the display will show the...

Страница 17: ...t above 25 C 78 F agitation and exposure to light must be avoided INTERFERENCES Interferences may be caused by strong oxidants 10 mL 2 2 Chlorine Dioxide 1 Add the content of one packet of HI 93738B 0...

Страница 18: ...g L Accuracy 5 g L 4 of reading Typical EMC 1 g L Deviation Light Source Tungsten lamp with narrow band interference filter 525 nm Method Adaptation of the ASTM Manual of Water and Environmental Tech...

Страница 19: ...g L Accuracy 1 g L 4 of reading Typical EMC 1 g L Deviation Light Source Tungsten lamp with narrow band interference filter 525 nm Method Adaptation of the ASTM Manual of Water and Environmental Techn...

Страница 20: ...5 00 mg L Resolution 0 01 mg L Accuracy 0 02 mg L 4 of reading Typical EMC 0 01 mg L Deviation Light Source Tungsten lamp with narrow band interference filter 575 nm Method Adaptation of the EPA meth...

Страница 21: ...0 to 1000 g L Resolution 1 g L Accuracy 10 g L 5 of reading Typical EMC 1 g L Deviation Light Source Tungsten lamp with narrow band interference filter 575 nm Method Adaptation of the EPA method The r...

Страница 22: ...sten lamp with narrow band interference filter 420 nm Method Adaptation of the ASTM Manual of Water and Environmental Technology method D1385 88 p Dimethylaminobenzaldehyde method The reaction between...

Страница 23: ...rference filter 525 nm Method Adaptation of the EPA Phenantroline method 315B for natural and treated waters The reaction between iron and reagents causes an orange tint in the sample REQUIRED REAGENT...

Страница 24: ...IONS Range 0 to 400 g L Resolution 1 g L Accuracy 10 g L 8 of reading Typical EMC 1 g L Deviation Light Source Tungsten lamp with narrow band interference filter 575 nm Method Adaptation of the TPTZ M...

Страница 25: ...ccessories see page 73 MEASUREMENT PROCEDURE Select the Molybdenum method using the procedure described in the Method Selection section see page 12 Fill the cuvette with 10 mL of unreacted sample up t...

Страница 26: ...r above 10 mg L Iron above 50 mg L Nickel above 50 mg L Nitrite as NO2 Sulfate above 200 mg L Highly buffered samples or with extreme pH may exceed the buffering capacity of the reagents Fill one grad...

Страница 27: ...amines Chloride above 100 ppm Chlorine above 2 ppm Copper Iron III Strong oxidizing and reducing substances Sulfide must be absent Nitrate SPECIFICATIONS Range 0 0 to 30 0 mg L Resolution 0 1 mg L Ac...

Страница 28: ...sten lamp with narrow band interference filter 575 nm Method Adaptation of the Ferrous Sulfate method The reaction between nitrite and the reagent causes a greenish brown tint in the sample REQUIRED R...

Страница 29: ...Nitrite LR SPECIFICATIONS Range 0 00 to 0 35 mg L Resolution 0 01 mg L Accuracy 0 02 mg L 4 of reading Typical EMC 0 01 mg L Deviation Light Source Tungsten lamp with narrow band interference filter...

Страница 30: ...eading Typical EMC 0 1 mg L Deviation Light Source Tungsten lamp with narrow band interference filter 420 nm Method Adaptation of the Standard Methods for the Examination of Water and Wastewater 18th...

Страница 31: ...scription Quantity HI 93710 0 Phenol Red Indicator 5 drops REAGENT SETS HI 93710 01 Reagents for 100 pH tests HI 93710 03 Reagents for 300 pH tests For other accessories see page 73 MEASUREMENT PROCED...

Страница 32: ...RANGE SPECIFICATIONS Range 0 0 to 30 0 mg L Resolution 0 1 mg L Accuracy 1 mg L 4 of reading Typical EMC Dev 0 1 mg L Light Source Tungsten lamp with narrow band interference filter 525 nm Method Ada...

Страница 33: ...cy 0 04 mg L 4 of reading Typical EMC Dev 0 01 mg L Light Source Tungsten lamp with narrow band interference filter 610 nm Method Adaptation of the Ascorbic Acid method The reaction between phosphate...

Страница 34: ...mg L Accuracy 0 03 mg L 3 of reading Typical EMC 0 01 mg L Deviation Light Source Tungsten lamp with narrow band interference filter 610 nm Method Adaptation of the ASTM Manual of Water and Environme...

Страница 35: ...HI 93731 01 Reagents for 100 tests HI 93731 03 Reagents for 300 tests For other accessories see page 73 MEASUREMENT PROCEDURE Select the Zinc method using the procedure described in the Method Select...

Страница 36: ...be caused by Aluminum above 6 mg L Cadmium above 0 5 mg L Copper above 5 mg L Iron above 7 mg L Manganese above 5 mg L Nickel above 5 mg L Zinc Place the cap and insert the cuvette into the instrumen...

Страница 37: ...eropolyBlue Zinc 0 00 to 3 00 mg L Zincon ACCESSORIES REAGENT SETS HI 93700 01 100 ammonia LR tests HI 93700 03 300 ammonia LR tests HI 93701 01 100 free chlorine tests powder HI 93701 03 300 free chl...

Страница 38: ...r Service Center nearest to you To find the Hanna Office in your vicinity check our home page at www hannainst com RecommendationsforUsers Before using these products make sure that they are entirely...

Страница 39: ...567 e mail sales hanna co jp Korea Tel 02 2278 5147 Fax 02 2264 1729 e mail mccoyhan chollian net Malaysia Tel 603 5638 9940 Fax 603 5638 9829 e mail hannamal tm net my Norway Tel 23 3811 00 Fax 23 38...

Отзывы: