background image

4

5

Where:

-log

 

I

/

I

o

=Absorbance (A)

I

o

=intensity of incident light beam

I

=intensity of light beam after absorption

ε

λ

=molar extinction coefficient at wavelength

 

λ

c =molar concentration of the substance
d =optical path through the substance

Therefore, the concentration "c" can be calculated from the absorbance of the substance as the other factors
are known.
Photometric chemical analysis is based on the possibility to develop an absorbing compound from a specific
chemical reaction between sample and reagents.
Given that the absorption of a compound strictly depends on the wavelength of the incident light beam, a
narrow spectral bandwidth should be selected as well as a proper central wavelength to optimize measurements.
The optical system of 

HI 83203

 is based on special subminiature tungsten lamps and narrow-band

interference filters to guarantee both high performance and reliable results.

Four measuring channels allow a wide range of tests.

Instrument block diagram (optical layout)

A microprocessor controlled special tungsten lamp emits radiation which is first optically conditioned and beamed
to the sample contained in the cuvette. The optical path is fixed by the diameter of the cuvette. Then the light

is spectrally filtered to a narrow spectral bandwidth, to obtain a light beam of intensity 

-I

o-

 or 

-I-.

The photoelectric cell collects the radiation 

-I-

 that is not absorbed by the sample and converts it into

an electric current, producing a potential in the mV range.
The microprocessor uses this potential to convert the incoming value into the desired measuring unit and to
display it on the LCD.
The measurement process is carried out in two phases: first the meter is zeroed and then the actual
measurement is performed.
The cuvette has a very important role because it is an optical element and thus requires particular attention. It
is important that both the measurement and the calibration (zeroing) cuvette are optically identical to provide the
same measurement conditions. Most of methods use the same cuvette  for both, so it is important that
measurements take place in the same optical point. The instrument and the cuvette cap have special marks
that must be aligned in order to obtain better 

reproducibility

.

The surface of the cuvette must be clean and not scratched. This is to avoid measurement interference due to
unwanted reflection and absorption of light. It is recommended not to touch the cuvette  walls with hands.
Furthermore, in order to maintain the same conditions during the zeroing and the measurement phases,
it is necessary to close the cuvette to prevent any contamination.

PRINCIPLE OF OPERATION

Absorption of Light is a typical phenomenon of interaction between electromagnetic radiation and matter.

When a light beam crosses a substance, some of the radiation may be absorbed by atoms, molecules or

crystal lattices.
If pure absorption occurs, the fraction of light absorbed depends both on the optical path length through the
matter and on the physical-chemical characteristics of substance according to the Lambert-Beer Law:

-log

 

I

/

I

o

 = 

ε

λ

 

c d

or

A  

ε

λ

 

c d

PRECISION AND ACCURACY

Precision is how closely repeated measurements agree

with each other. Precision is usually expressed as

standard deviation (SD).
Accuracy is defined as the nearness of a test result to

the true value.
Although good precision suggests good accuracy, precise

results can be inaccurate. The figure explains these

definitions.
For each method, the precision is expressed in the

related measurement section as standard deviation

at a specific concentration value of the analite. The

standard deviation is obtained with a single instrument
using a representative lot of reagents.

SPECIFICATIONS

Light Life

Life of the instrument

Light Detector

Silicon Photocell

Environment

0 to 50°C (32 to 122°F);
max 90% RH non-condensing

Power Supply

external 12 Vdc power adapter
built-in rechargeable battery

Dimensions

235 x 200 x 110 mm (9.2 x 7.87 x 4.33")

Weight

0.9 Kg

For specifications related to each method (e.g. range, precision, etc.) refer to the related measurement

section.

Содержание HI 83203

Страница 1: ...1 Instruction Manual HI 83203 Multiparameter Bench Photometer for Aquaculture w w w h a n n a i n s t c o m This Instrument is in Compliance with the CE Directives...

Страница 2: ...tometer can be connected to a PC via an USB cable The optional HI 92000 Windows Compatible Software helps users manage all their results Dear Customer Thank you for choosing a Hanna product Please rea...

Страница 3: ...are optically identical to provide the same measurement conditions Most of methods use the same cuvette for both so it is important that measurements take place in the same optical point The instrumen...

Страница 4: ...creen NEED TO KNOW HI 83203 has a powerful interactive user support that assists the user during the analysis process Each step in the measurement process is help supported A tutorial mode is availabl...

Страница 5: ...a b a c a b shake gently shakevigorously b c USING CUVETTES Proper mixing of the cuvette is done by shaking the cuvette moving the cuvette up and down The movement may be gentle or vigorous This mixi...

Страница 6: ...before performing tests Safety equipment Wear suitable eye protection and clothing when required and follow instructions carefully Reagent spills If a reagent spill occurs wipe up immediately and rins...

Страница 7: ...tothe setup menu without saving the new value SETUP SETUP SETUP SETUP SETUP DATA MANAGEMENT The instrument features a data log function to help you keep track of all your analysis The data log can hol...

Страница 8: ...ument serial number Press ESC to return to the Setup mode HELP MODE HELP MODE HELP MODE HELP MODE HELP MODE HELP MODE Date Time This option is used to set the instrument s date and time Press Modify f...

Страница 9: ...arious aliphatic and aromatic amines Ammonia MR x 4 x 4 Ammonia MR SPECIFICATIONS Range 0 00 to 10 00 mg L Resolution 0 01 mg L Accuracy 0 05 mg L 5 of reading Typical EMC 0 01 mg L Deviation Light So...

Страница 10: ...amines sulfide various aliphatic and aromatic amines x 4 x 4 Ammonia LR SPECIFICATIONS Range 0 00 to 3 00 mg L Resolution 0 01 mg L Accuracy 0 04 mg L 4 of reading Typical EMC 0 01 mg L Deviation Ligh...

Страница 11: ...ke gently again x 3 x 3 FREE CHLORINE FREE CHLORINE FREE CHLORINE FREE CHLORINE FREE CHLORINE SPECIFICATIONS Range 0 00 to 2 50 mg L Resolution 0 01 mg L Accuracy 0 03 mg L 3 of reading Typical EMC 0...

Страница 12: ...ests powder For other accessories see page 46 MEASUREMENT PROCEDURE Select the Total Chlorine method using the proce dure described in the Method Selection section see page 11 Fill the cuvette with 10...

Страница 13: ...partially or may rapidly fade To resolve this neutralize the sample with diluted HCl or NaOH Total Chlorine Total Chlorine Press ZERO key The meter will show 0 0 when the meter is zeroed and ready for...

Страница 14: ...5 00 mg L Resolution 0 01 mg L Accuracy 0 02 mg L 4 of reading Typical EMC 0 01 mg L Deviation Light Source Tungsten lamp with narrow band interference filter 575 nm Method Adaptation of the EPA meth...

Страница 15: ...0 to 1000 g L Resolution 1 g L Accuracy 10 g L 5 of reading Typical EMC 1 g L Deviation Light Source Tungsten lamp with narrow band interference filter 575 nm Method Adaptation of the EPA method The r...

Страница 16: ...amines Chloride above 100 ppm Chlorine above 2 ppm Copper Iron III Strong oxidizing and reducing substances Sulfide must be absent Nitrate SPECIFICATIONS Range 0 0 to 30 0 mg L Resolution 0 1 mg L Ac...

Страница 17: ...sten lamp with narrow band interference filter 575 nm Method Adaptation of the Ferrous Sulfate method The reaction between nitrite and the reagent causes a greenish brown tint in the sample REQUIRED R...

Страница 18: ...Nitrite LR SPECIFICATIONS Range 0 00 to 0 35 mg L Resolution 0 01 mg L Accuracy 0 02 mg L 4 of reading Typical EMC 0 01 mg L Deviation Light Source Tungsten lamp with narrow band interference filter...

Страница 19: ...ading Typical EMC 0 1 mg L Deviation Light Source Tungsten lamp with narrow band interference filter 420 nm Method Adaptation of the Standard Methods for the Examination of Water and Wastewater 18th e...

Страница 20: ...scription Quantity HI 93710 0 Phenol Red Indicator 5 drops REAGENT SETS HI 93710 01 Reagents for 100 pH tests HI 93710 03 Reagents for 300 pH tests For other accessories see page 46 MEASUREMENT PROCED...

Страница 21: ...RANGE SPECIFICATIONS Range 0 0 to 30 0 mg L Resolution 0 1 mg L Accuracy 1 mg L 4 of reading Typical EMC Dev 0 1 mg L Light Source Tungsten lamp with narrow band interference filter 525 nm Method Ada...

Страница 22: ...cy 0 04 mg L 4 of reading Typical EMC Dev 0 01 mg L Light Source Tungsten lamp with narrow band interference filter 610 nm Method Adaptation of the Ascorbic Acid method The reaction between phosphate...

Страница 23: ...ETHODS The analyzed data can be managed using Hanna s product HI 92000 Windows Compatible Software DATA MANAGEMENT Theinstrumentshowsclearwarningmessageswhenerroneousconditionsappearandwhenmeasuredval...

Страница 24: ...instruments may cause unacceptable interferences to other electronic equipments this requiring the operator to take all necessary steps to correct interferences Any variation introduced by the user to...

Страница 25: ...567 e mail sales hanna co jp Korea Tel 02 2278 5147 Fax 02 2264 1729 e mail mccoyhan chollian net Malaysia Tel 603 5638 9940 Fax 603 5638 9829 e mail hannamal tm net my Norway Tel 23 3811 00 Fax 23 38...

Отзывы: