Hanna Instruments HI 758 Скачать руководство пользователя страница 2

Reagent Sets

HI758-26

reagent set for 25 marine calcium tests

Other Accessories

HI758-11

marine calcium certified standard kit

HI731315

Glass Cuvettes and Caps for Checker

®

 HC Colorimeters (2 pcs.)

HI731318

cloth for wiping cuvettes (4 pcs.)

HI731339P

0.1 mL minipipette

HI731349P

tip for 0.1 mL minipipette (10 pcs.)

HI740028P

1.5V AAA batteries (12 pcs.)

HI70436M

deionized water (230 mL)

HI93703-50

cuvette cleaning solution (230 mL)

Accessories

To conserve battery, the instrument shuts down after 
10 minutes of non-use. One fresh battery lasts for a minimum of 
5000 measurements. When the battery is dead the instrument 
will display “bAd” then “bAt” for 1 second and then turns off. To 
replace the battery, turn the instrument off by holding the button 
until the meter shuts off. Turn the instrument upside down and 
remove the battery cover with a screwdriver. Remove the battery 
from its location and replace it with a new one, inserting the 
negative end first. Replace the battery cover and screw.

It is important that the sample does not contain any debris. 
Whenever the cuvette is placed into the measurement cell, it must 
be dry outside, and completely free of fingerprints, oil and dirt. Wipe 
it thoroughly with 

HI731318

 or a lint-free cloth prior to insertion.

Shaking the cuvette can generate bubbles, causing higher readings. 
To obtain accurate measurements, remove bubbles by swirling or 
by gently tapping the cuvette. Do not let the reacted sample stand 
for too long after reagent is added, as accuracy will be affected. 
After the reading, it is important to immediately discard the sample, 
otherwise the glass might become permanently stained. 

• Attach the pipette tip.
• Press the button down to the 

first stop

.

• Immerse the pipette tip in the liquid approximately 2-3 mm.
• Slowly let the button move back to the 

original position

, wait 2 seconds.

• Remove the pipette tip from the liquid.
• To dispense the liquid, place the pipette tip on the inside wall of the container.
• Slowly press the button down to the 

first stop

.

• Wait until all of the liquid has been dispensed.
• Press the button down to the 

second stop

, this will allow any remaining liquid to be dis-

pensed.

Turn the meter on by pressing the button. All 
segments will be displayed. When the display 
shows 

“Add”

“C1”

 with 

“Press”

 blinking, the 

meter is ready.

Add 1 mL of 

HI7581

 Reagent A to the cuvette 

using the 1 mL syringe. Use the plastic pipette 
to fill the cuvette to the 10 mL mark with 
deionized water and replace the cap. Invert 
3-5 times to mix.

Place the cuvette into the meter and close 
the meter’s cap. Press the button. When 
the display shows 

“Add”

“C2”

 with 

“Press”

 

blinking the meter is zeroed. 

Remove the cuvette from the meter and 
unscrew the cap. Use the minipipette to add 
0.1 mL of sample to the cuvette. 

Add 1 packet of 

HI7582

 Reagent B. 

Replace the cap and shake vigorously for 
15 seconds or until the powder is completely 
dissolved. Allow air bubbles to dissipate for 
15 seconds before taking a reading. 

Place the cuvette back into the meter and 
close  the meter’s cap. Press the button. 
The meter displays concentration in ppm of 
calcium. The meter automatically turns off 
after 10 minutes.

10 

mL

First stop Second stop

Original position

Measurement Procedure

Minipipette Instructions:

Tips for an Accurate Measurement

Battery Management

IST758   04/17

HI758

 is warranted for a period of one year after date of purchase against defects in 

workmanship and materials when used for their intended purpose and maintained 
according to instructions. This warranty is limited to repair or replacement free of charge. 
Damage due to accidents, misuse, tampering or lack of prescribed maintenance is not 
covered. If service is required, contact your local Hanna Instruments Office. If under 
warranty, report the model number, date of purchase, serial number and the nature of the 
problem. If the repair is not covered by the warranty, you will be notified of the charges 
incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Return 
Goods Authorization number from the Technical Service department and then send it with 
shipping costs prepaid. When shipping any instrument, make sure it is properly packed 
for complete protection.

Before using Hanna Instruments products, make sure that they are entirely suitable for 
your specific application and for the environment in which they are used. Operation of 
these instruments may cause unacceptable interferences to other electronic equipment, 
thus requiring the operator to take all necessary steps to correct such interferences. Any 
variation introduced by the user to the supplied equipment may degrade the instrument’s 
EMC performance. To avoid damages or burns, do not put the instrument in microwave 
oven. For yours and the instrument safety do not use or store the instrument in hazardous 
environments.

Recommendations for Users

Warranty

Hanna Instruments reserves the right to modify the design, 

construction, or appearance of its products without advance notice.

Отзывы: