V.   Equipment required

Equipment required

Equipment required

Equipment required

Equipment required:

Hanna HI 5315 double junction reference electrode
with HI 7075 fill solution for measurements with

        HI 4010.

Hanna HI 4222 pH/ISE/mV meter or other suitable
ion or pH/mV meter. (Note: log/linear graph paper is
useful if an ISE (ion) meter is not available.

Hanna HI 180 magnetic stirrer or equivalent with
stirring bars (HI 731320).  (Note: isolate beakers from
stirrer motor heat by placing insulating material such
as foam or cork between them).

Hanna HI 76404 electrode holder or equivalent.

Plastic beakers (HI 740036P) or other suitable mea-
surement vessel.

VI.  

VI.  

VI.  

VI.  

VI.  Solutions Required

Solutions Required

Solutions Required

Solutions Required

Solutions Required

Standard for Fluoride Measurements
Select appropriate Hanna Instruments standard and ISA
from the list below:

Part number

0.1 M Sodium Fluoride, 500 mL

HI 4010-01

100 ppm, 500 mL

HI 4010-02

1000 ppm, 500 mL

HI 4010-03

10 ppm with TISAB II, 500 mL

HI 4010-10

1 ppm with TISAB II, 500 mL

HI 4010-11

2 ppm with TISAB II, 500 mL

HI 4010-12

ISA
TISAB II, 500 mL

HI 4010-00

TISAB II, 1 gallon

HI 4010-05

TISAB III , 500 mL

HI 4010-06

KIT:

HI 4010-30

Contains four 500 mL bottles of:
TISAB II

HI 4010-00

10 ppm with TISAB II

HI 4010-10

1 ppm with TISAB II

HI 4010-11

IV.  

IV.  

IV.  

IV.  

IV.  Design elements of the HI 4010 and HI 4110

Design elements of the HI 4010 and HI 4110

Design elements of the HI 4010 and HI 4110

Design elements of the HI 4010 and HI 4110

Design elements of the HI 4010 and HI 4110

electrodes

electrodes

electrodes

electrodes

electrodes

Upper Cap

Spring

O-Ring

Ceramic Junction
on Inner Stem

Fill Hole

O-Ring and
Plug

Sensing
Crystal

Outer Sleeve

Upper
Threads

Liquid junction

FLUO

RIDE

HI 4010

Cap

Sensing
Crystal

Sensor
Handle

*

*

*

*

*

Standards that are asterisked contain TIASB II and are
ready to use withour further additions.

*

Содержание HI 4110

Страница 1: ...Instruction Manual HI 4110 FLUORIDE COMBINATION FLUORIDE HI 4010 HI 4010 HI 4010 HI 4010 HI 4010 HI 4010 HI 4110 HI 4110 HI 4110 HI 4110 HI 4110 Fluoride Ion Selective Electrode Half cell Combination...

Страница 2: ...o forms HF species below pH 5 Increase pH in these cases above 5 for a total fluoride measurement Operating Temperature 0 80 C Operating pH 5 to 8 pH Dimensions 12 mm OD X 120 mm nominal insertion 0 4...

Страница 3: ...standard and ISA from the list below Part number 0 1 M Sodium Fluoride 500 mL HI4010 01 100 ppm 500 mL HI4010 02 1000 ppm 500 mL HI4010 03 10 ppm with TISAB II 500 mL HI4010 10 1 ppm with TISAB II 500...

Страница 4: ...ntact Hanna Instruments for guidelines with other applications VII VII VII VII VII General Guidelines General Guidelines General Guidelines General Guidelines General Guidelines Calibration standards...

Страница 5: ...r is inverted an air pocket will be seen which is normal 3 Prepare HI 5315 reference electrode by filling outer electrolytereservoirwithHI7072 4 Place sensor and reference electrodes into electrode ho...

Страница 6: ...uoridestandardfor1hour Repeatsection IX If the reading is jumpy or unstable shake sensor down seesectionVIII If the membrane is damaged the response becomes extremely sluggish or the slope of the elec...

Страница 7: ...ution to condition before calibration XI XI XI XI XI Direct Calibration and Measurement Direct Calibration and Measurement Direct Calibration and Measurement Direct Calibration and Measurement Direct...

Страница 8: ...and Care of the HI 4010 and Storage and Care of the HI 4010 and Storage and Care of the HI 4010 and Storage and Care of the HI 4010 and HI 4110 sensors HI 4110 sensors HI 4110 sensors HI 4110 sensors...

Страница 9: ...r 6 monthsfromdateofpurchasewhenusedfortheirintended purpose and maintained according to instructions If they failtoworkwhenfirstusedcontactyourdealerimmediately Damageduetoaccidents misuse misapplica...

Отзывы: