V.  Equipment required:

V.  Equipment required:

V.  Equipment required:

V.  Equipment required:

V.  Equipment required:

Hanna HI 5315 Double Junction Reference Electrode
with HI 7072 Fill Solution for use with HI 4015.

Hanna HI 4222 pH/ISE/mV meter or other suitable
ion or pH/mV meter. (Note: log/linear graph paper is
useful if an ISE (ion) meter is not available).

Hanna HI 180 Magnetic Stirrer or equivalent with
Teflon coated stirring bars (HI 731320).  (Note: iso-
late beakers from stirrer motor heat by placing insu-
lating material such as foam or cork between them).

Hanna HI 76404 Electrode Holder or equivalent.

Plastic beakers (HI 740036P) or other suitable mea-
surement vessel.

VI. 

VI. 

VI. 

VI. 

VI. Solutions Required for Calibration

Solutions Required for Calibration

Solutions Required for Calibration

Solutions Required for Calibration

Solutions Required for Calibration

Select appropriate Hanna Instruments standard and ISA
from the list below:

0.1 M Silver Standard, 500 mL

HI 4015-01

ISA for Silver measurement, 500 mL

HI 4000-00

SAOB for Sulfide measurement, 500 mL HI 4015-00
Note: For Sulfide standard, see directions that follow.

Silver Standards
For Molar solutions:
Using volumetric pipettes and glassware make serial dilu-
tions of HI 4015-01  to approximately bracket the concen-
tration of the samples.  Standards with concentrations

<

 10

-3 

M should be prepared daily. Store in dark plastic

bottles.
Two mL of Hanna ISA for Halide electrodes (HI 4000-00)
should be added to 100 mL of sample or standard.
For ppm solutions:
Prepare 1078.7 ppm Silver standard by diluting HI 4015-
01: Pipette 100 mL standard to a 1 liter volumetric flask.
Add deionized water to volume. Using additional  pipettes
and glassware make serial dilutions of this 1078.7 ppm
standard to bracket the sample concentra-

SIL

VER/SULFIDE

HI 4015

Cap

Sensing
Membrane

Sensor
Handle

Upper Cap

Spring

O-Ring

Ceramic Junction
on Inner Stem

Fill Hole

O-Ring and
Plug

Sensing
Membrane

Outer Sleeve

Upper
Threads

Liquid junction

Содержание HI 4015

Страница 1: ...HI 4015 HI 4015 HI 4015 HI 4015 HI 4115 HI 4115 HI 4115 HI 4115 HI 4115 Silver Sulfide Ion Selective Electrode Instruction Manual Half cell Combination SILVER SULFIDE HI 4015 HI 4115 SILVER SULFIDE CO...

Страница 2: ...rating pH 2 8 pH For Silver 12 14 pH For Sulfide Dimensions 12 mm OD X 120 mm nominal insertion 0 47 X4 72 Connection BNC III III III III III Theory of Operation Theory of Operation Theory of Operatio...

Страница 3: ...Calibration Solutions Required for Calibration Solutions Required for Calibration Solutions Required for Calibration Select appropriate Hanna Instruments standard and ISA from the list below 0 1 M Sil...

Страница 4: ...M Lead Perchlorate standard Pb ClO4 2 Pb2 S2 PbS Use the electrode pair as an indicator for S2 to determine the endpoint See Section XII V V V V Vs s s s s Volume of the standard sulfide solution use...

Страница 5: ...dd a few drops HI 7072 fill solution to the electrode wetting the o ring and rinsing out the fill solution chamber De ion ize d W at er Remove Parafilm VIII VIII VIII VIII VIII Electrode Preparation E...

Страница 6: ...ly sluggish or the slope of the electrode has decreased significantly and procedures above have not helped the sensor should be replaced IX IX IX IX IX Quick Check of Electrode Slope Quick Check of El...

Страница 7: ...ith two or more freshly made standards that are in the linear measurement range of the unknowns Unknowns are read directly At very low levels of silver special precautions must be em ployed for reprod...

Страница 8: ...ardVstandard Vsample VS Vsample Vstandard VISA VT Vsample VISA VS Analyte Subtraction S2 This method is based on a titration reaction with the electrode responsing to a reagent that reacts with the sa...

Страница 9: ...w mV mV2 mV1 S is the slope of the silver electrode 3 The unknown sulfide concentration in the sample Csam can then be determined by the following equation For an example a sensor is placed into 100 m...

Страница 10: ...015 and Storage and Care of the HI 4015 and Storage and Care of the HI 4015 and Storage and Care of the HI 4015 and Storage and Care of the HI 4015 and HI 4115 sensors HI 4115 sensors HI 4115 sensors...

Страница 11: ...6 monthsfromdateofpurchasewhenusedfortheirintended purpose and maintained according to instructions If they failtoworkwhenfirstusedcontactyourdealerimmediately Damageduetoaccidents misuse misapplicat...

Отзывы: