IX.   Quick Check of Electrode Slope

Quick Check of Electrode Slope

Quick Check of Electrode Slope

Quick Check of Electrode Slope

Quick Check of Electrode Slope

Connect sensors to pH/mV/ISE meter

Place meter in mV mode.

Place 100 mL of DIW into a beaker with stir bar.

Place reference and measuring half-cell into prepared
sample.

Add 1 mL of a standard (0.1 M or 1000 ppm stan-
dard) to beaker. Record the mV value when stable.

Add an additional 10-mL of standard to the solution.
Record the mV when reading has stabilized.  This
value should be less than the previous noted (more
negative).

Determine the difference between the two-mV values.
An acceptable value for this slope is -56 ± 4 mV.

X.   

X.   

X.   

X.   

X.   Corrective action

Corrective action

Corrective action

Corrective action

Corrective action

Verify protective cap has been removed.(HI 4007)

Verify plastic film has been removed from inner
stem.(HI 4107)

Verify electrodes are connected properly to meter and
meter is powered.

Verify dilute standards are freshly made and stored.
Remake solutions if appropriate.

If the sensor slope just misses the suggested slope
window, soaking the sensor in a dilute standard may
solve the problem. (Choose 10

-3 

M chloride or 100

ppm standard).

A scratched sensing surface can be polished with HI
4000-70 polishing strip.  Cut off approximately 1
inch of the micro-abrasive strip. Wet the frosted side
with deionized water 

 

and place against damaged

membrane of the electrode.  Place your thumb against
the shiny backing  and slowly rotate back and forth
while applying gentle pressure.  Continue polishing
until you are satisfied with the surface.

Содержание HI 4007

Страница 1: ...HI 4007 HI 4007 HI 4007 HI 4007 HI 4007 HI 4107 HI 4107 HI 4107 HI 4107 HI 4107 Chloride Ion Selective Electrode Instruction Manual CHLORIDE HI 4007 HI 410 COMBINATION 7 CHLORIDE Half cell Combinatio...

Страница 2: ...utions The HI 4007 is a solid state half cell sensor that requires a separate reference The HI 4107 is a combination ion selective electrode II II II II II Specifications Specifications Specifications...

Страница 3: ...asareferenceelectrode incorporated in its design The silver chloride pellet is prac tically insoluble in the test solutions being measured and producesapotentialchangeduetochangesinthesample s ion act...

Страница 4: ...4107 Design elements of the HI 4007 and HI 4107 Design elements of the HI 4007 and HI 4107 electrodes electrodes electrodes electrodes electrodes CHLORIDE HI 4007 Cap Sensing Membrane Sensor Handle U...

Страница 5: ...lutions Required Solutions Required for Chloride Measurements for Chloride Measurements for Chloride Measurements for Chloride Measurements for Chloride Measurements Select appropriate Hanna Instrumen...

Страница 6: ...libration standards and sample solutions should have the same ionic strength ISA acetate buffer or oxidizing reagent should be added to both samples and standards in the same ratio Calibration standar...

Страница 7: ...ing normal use fill solution will slowly drain out ofthetaperedconejunctionatthelowerportionofthe electrode Excessiveloss 4cmdropwithin24hours isnotnormal Ifthisoccursverifycapistightenedand the inter...

Страница 8: ...m seal found over ceramic junction on inner stem and discard This is only used for ship ping and long term storage 2 Rinseinnerstemwithdeionizedwatermakingcertain to wet the o ring found on the inner...

Страница 9: ...verify elec trodereturnstoitsoriginalposition Youmayneedto gently assist for this to occur 7 Tighten the electrode cap onto the body and fill elec trode body until fill solution volume is just below f...

Страница 10: ...action Corrective action Corrective action Corrective action Corrective action Verify protective cap has been removed HI 4007 Verify plastic film has been removed from inner stem HI4107 Verify electr...

Страница 11: ...he meter is calibrated with two or more freshly made stan dards that are in the linear measurement range of the unknowns Morecalibrationstandardsarerequiredinnon linear regions Unknowns are read direc...

Страница 12: ...ements using the oxidizing reagent must be made 10 minutes after mixing and not permitted react longer 3 Follow sectionVII GeneralGuidelinestooptimizetest set up 4 Duringcalibrationitisbesttostartwith...

Страница 13: ...mplifies the method greatly Example Chloride ion determination with known addition 1 A 50 mL sample of unknown Vsample is placed in a clean plastic beaker with a chloride sensor Add 2 mLofacetatepHbuf...

Страница 14: ...r See Section VI Titration Titration Titration Titration Titration XIV S XIV S XIV S XIV S XIV Storage and Care of the HI 4007 and torage and Care of the HI 4007 and torage and Care of the HI 4007 and...

Страница 15: ...s Conversion tables Conversion tables Conversion tables Conversion tables For Cl For Cl For Cl For Cl For Cl Multiply by Multiply by Multiply by Multiply by Multiply by Moles L M to ppm mg L 3 500x104...

Страница 16: ...6 monthsfromdateofpurchasewhenusedfortheirintended purpose and maintained according to instructions If they failtoworkwhenfirstusedcontactyourdealerimmediately Damageduetoaccidents misuse misapplicat...

Отзывы: